Voltage Setting For Replacing The Transformer - Toshiba Aplio SSA-770A Service Manual

Diagnostic ultrasound system
Table of Contents

Advertisement

No. 2D730-148E*O
ccTrans-unit>>
Circuit protector for
monitor AC output
LED for error display
~
P41 monitor AC output
j~p
J52 relay control input
TRS LED control input
J51 AC output for AC/DC tray
Circuit protector for OEM AC output
ZAUTION:
Voltage setting for replacing the transformer
tray and TRANS-unit
The switches used to select the output voltage according to the input voltage are installed
on the transformer
tray and the TRANS-unit.
When the switches of the replacement
units (transformer
tray and TRANS-unit)
are not set
correctly, the power supply may not start up correctly, or a malfunction
of the connection
equipment
may result. Be sure to check the settings.
For the transformer
tray
100 to 120 VAC
aIll;:
"115" must be displayed on Sl and S3.
The direction of S2 must be the same as that of Sl.
230 VAC
czll~~
"230" must be displayed on Sl and S3.
The direction of S2 must be the same as that of 53.
For the TRANS-unit
100 to 120 VAC
115
'D
El
"115" must be displayed.
115
kz
"115" must be displayed.
z
The direction must be the same as that of Sl and 52.
230 VAC
"230" must be displayed.
"230" must be displayed.
The direction must be the same as that of Sl and 52.
6-4