Panel Switch Layout And Switch Functions After Configuration - Toshiba Aplio SSA-770A Service Manual

Diagnostic ultrasound system
Table of Contents

Advertisement

No. 2D730-148E*O
15.3
Panel Switch Layout and Switch Functions After Configuration
Change
Replace the key tops as shown in the figure according to the selected layout.
Confirm that the configuration
has been changed correctly.
If the configuration
has not been
changed correctly, correct it.
(1)
TMSJ abdomen A (TMSJ-Abd-A.bat)
(Pattern A)
The functions of the seven locations indicated in the figure below are changed.
For cl > and c2>, replace the elliptic label below the switch. For c3> to c7>, replace the
key tops.
No.
Switch after
layout change
Function
Cl>
1 2D FREQ
1 Same as the 2D FREQ switch before the layout change
<2>
2D FOCUS
Same as the 2D FOCUS switch before the layout change
<3>
UPDATE
Same as the UPDATE switch before the layout change
<4>
1 FULL
1 Same as the FULL switch before the layout change
<5>
1 DUAL
1 Same as the DUAL switch before the layout change
<6>
1 STORE
1 Same as the STORE switch before the layout change
<7>
1 FREEZE
1 Same as the FREEZE switch before the layout change
15-2