Status Led; Starting Up The Adjustment Guide File And Adjusting Image Quality - Toshiba Aplio SSA-770A Service Manual

Diagnostic ultrasound system
Table of Contents

Advertisement

No. 2D730-148E*O
(3)
Status LED
The LED at the lower front of the LCD monitor indicates the monitor status.
(4)
LED indication
I
Status
OFF
Blinking green (ON/OFF at 2-s intervals)
AC power OFF
Forced power save mode
Lights green
1 Startup, display ON
Lights orange
Blinking orange (ON/OFF at 2-s intervals)
Automatic power save mode (No signal)
Signal error
Blinking orange (ON/OFF at l-s intervals)
Automatic power save mode
(Abnormal temperature
inside the LCD
monitor)
Starting up the adjustment
guide file and adjusting image quality
(a)
Double-click "C:\LCD\Loadadj.exe".
(b)
Confirm that the Auto Adjustment
guide screen is started up.
(c)
Hold down both the [Y] switch and the [+] switch on the front of the LCD monitor for 3
seconds to display the service mode menu (main menu).
7-11

Advertisement

Table of Contents
loading

This manual is also suitable for:

Aplio ssa-700a

Table of Contents