Log off or exit UNICORN
Follow the instruction below to log off or exit UNICORN. This can be performed from any
of the UNICORN modules.
If you want to...
log off UNICORN
Getting Started with ÄKTAavant and UNICORN 6 28-9440-69 AB
Action
Prepare at least 10 ml storage solution by mixing equal volumes of a
standard buffer pH 4 and a 1 M Potassium Nitrate (KNO
Fill a syringe with approximately 10 ml of the storage solution. Connect the
syringe to the pH valve port Cal, and inject the storage solution.
Press
.
Result: The pH valve switches back to the default position and the Calibration
dialog closes. No calibration is performed.
then...
select File:Log off.
Result: All open UNICORN modules close and the Log On
dialog opens.
8 Run a method
8.5 After run procedures
) solution.
3
113