GE C-PT Installation Manual page 53

Ultrasonic flow transducer
Table of Contents

Advertisement

6.1b Mounting the TMCF in a 180
4. Position the blocks in a common plane of the pipe, but not on the top or bottom. The blocks should be opposite
o
each other (180
).
5. Wrap the remaining chains around the pipe and fasten them to the screw hooks on the other block. Make sure the
chains on either side of the pipe are approximately the same length.
6. Tighten and/or loosen the chains using the screw hooks on both blocks to fine tune the angle between the blocks.
Make sure the chains are not twisted or slanted.
7. Proceed to Mounting the Transducers into the TransFlection Mode Clamping Fixture - TMCF on page 47.
C-PT Ultrasonic Flow Transducer Installation Guide
o
Configuration (cont.)
End View
End View
End View
45