5.4d Mounting Transducers into the MCF (cont.)
4. Place the transducers in the appropriate blocks. Make sure the transducers are oriented as shown below.
CAUTION FOR ATEX CERTIFIED INSTALLATIONS
To ensure against impact to the transducer face, it must always be mounted flush to the pipe.
Note: If the transducer cables are already connected, you must determine the upstream and downstream directions of
the pipe and place the transducers into the appropriate blocks.
Downstream Transducer
Flow
Top View
5. Use the pressure bolt to secure the transducer in place. The pressure bolt should fit into the dimple. Hand-tighten
enough to hold the transducer in place. Do not overtighten so that the fixture lifts off the pipe.
WARNING! Do not tighten the pressure bolt so that the magnet separates from the pipe.
Pressure Bolt
Counter Nut
Top View