Chapter 5. Using the Configuration Menu
5.4.4 Entering Saturation Constants (cont.)
4. Select the Chemical and enter its percentage of the overall composition.
5. Repeat steps 2-4 until all of the data has been entered. After you select Save , the Cs Table will appear.
Note: At least 90% of the total composition must be entered. For any compositions that total less than 100%,
each chemical will be extrapolated to bring the total to 100%.
5.4.4.1 Entering Saturation Constants Manually
To manually enter the Cs values for your application, complete the following steps:
1. Select Manual Entry .
2. Tap the Channel button and specify the channel for which the Cs curve should be applied.
3. Use the Insert Row and Remove Row buttons to configure the table with up to 10 rows.
4. To enter or edit the data in the table, tap the desired table cell to highlight it in blue. Then, tap the Edit
Value button to enter the data for that cell. Repeat this process until all of the data has been entered.
Note: The Copy and Paste buttons can be used to copy data from another table into this new table.
5. After all of the data has been entered, tap the Sort button to be sure the data points are arranged in
ascending order of Temp C values. Then, tap the Check button to save the table and tap the Return
button to return to the Configuration Menu.
74
Figure 66: Entering Constants Manually
moisture.IQ User's Manual