Connecting Alarms - GE moisture.iq User Manual

Multi-channel moisture and oxygen analyzer
Hide thumbs Also See for moisture.iq:
Table of Contents

Advertisement

Chapter 1. Installation

1.13 Connecting Alarms

Note: Customers must provide their own cables for connecting alarm relays. Acceptable cables range from 12 to
24 AWG.
The moisture.IQ has two high/low alarm relays per channel. There is also one system fault alarm relay, which is
not channel specific. Each alarm relay has single-pole, double-throw contacts that consist of the following:
Normally Open (NO)
Common (C)
Normally Closed (NC)
Table 6 and Figure 26 below list alarm wiring connections.
Description
Normally Open
Normally Closed
1
NO
30
Table 6: Pin Designations for Relay Contacts
Fault
Common
Alarms
2
3
5
4
C
NC
RTN
RTN
Alarm A
Figure 26: Alarm Wiring
Alarm A
1
1
2
2
3
3
6
7
8
C
NO
NC
Alarm B
Alarm B
6
7
8
Fault Alarm Connector
1
2
3
C
NO
NC
moisture.IQ User's Manual

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents