Multi-channel moisture and oxygen analyzer
Hide thumbs Also See for moisture.iq:
Table of Contents

Advertisement

Function Parameter
Channel 6 /
Output A
Channel 6 /
Output B
3000
All Alarm
Status
Channel 1 /
Alarm A
moisture.IQ User's Manual
Table 13: Modbus Register Map (cont.)
Sub
Modbus
Parameter
ID
Address
Output
600 2612
(percent of
scale)
State
602 2602
Units
604 2604
Type
606 2606
Zero
608 2608
Span
610 2610
Output
620 2632
(percent of
scale)
State
622 2622
Units
624 2624
Type
626 2626
Zero
628 2628
Span
630 2630
0
3012
Status
100 3100
State
102 3102
Units
104 3104
Type
Access
32-bit float
RW
Read
Holding
(0x03)
32-bit Integer RW
Read
Holding
(0x03)
32-bit Integer RW
Read
Holding
(0x03)
32-bit Integer RW
Read
Holding
(0x03)
32-bit float
RW
Read
Holding
(0x03)
32-bit float
RW
Read
Holding
(0x03)
32-bit float
RW
Read
Holding
(0x03)
32-bit Integer RW
Read
Holding
(0x03)
32-bit Integer RW
Read
Holding
(0x03)
32-bit Integer RW
Read
Holding
(0x03)
32-bit float
RW
Read
Holding
(0x03)
32-bit float
RW
Read
Holding
(0x03)
32-bit Integer RO
Read Input
(0x04)
32-bit Integer RO
Read Input
(0x04)
32-bit Integer RO
Read
Holding
(0x04)
32-bit Integer RW
Read
Holding
(0x03)
Appendix D. Modbus Register Map
Modbus
Qty. of
Request
Registers
2
When TEST MODE
enabled, force
output to value
2
2 = TEST MODE,
All else = run mode
2
Measurement/Unit
Code 0 = off,
others See tab
2
1 = 4-20 mA,
2 = 0-20 mA,
3 = 0-2 volts
2
2
2
When TEST MODE
enabled, force
output to value
2
2 = TEST MODE,
All else = run mode
2
Measurement/Unit
Code 0 = off,
others see tab
2
1 = 4-20 mA,
2 = 0-20 mA,
3 = 0-2 volts
2
2
2
Bit wise Alarm
Status. 1 =
Energized
0 = De-energized
2
1 = Energized
0 = De-energized
2
0 = Disabled,
1 = Enabled
2
Measurement/Unit
Code 0 = off,
others see tab
Notes
131