Appendix D. Modbus Register Map - GE moisture.iq User Manual

Multi-channel moisture and oxygen analyzer
Hide thumbs Also See for moisture.iq:
Table of Contents

Advertisement

Appendix D. Modbus Register Map

Note: The Modbus Register Map in Table 13 below has some fields in gray boxes. These are currently not
supported and will return a value of –1.0.
Function Parameter
0
Error register,
non-latching.
Read only.
1000
Latching Error
Register
2000
Analog Output
(4-20)
Channel 1 /
Output A
Channel 1 /
Output B
moisture.IQ User's Manual
Table 13: Modbus Register Map
Sub
Modbus
Parameter
ID
Address
0
0
0
1000
0
Output
100 2112
(percent of
scale)
State
102 2102
Units
104 2104
Type
106 2106
Zero
108 2108
Span
110 2110
Output
120 2132
(percent of
scale)
State
122 2122
Units
124 2124
Type
126 2126
Zero
128 2128
Span
130 2130
Type
Access
32-bit Integer RO
Read Input
(0x04)
32-bit Integer RW
Read
Holding
(0x03)
32-bit float
RW
Read
Holding
(0x03)
32-bit Integer RW
Read
Holding
(0x03)
32-bit Integer RW
Read
Holding
(0x03)
32-bit Integer RW
Read
Holding
(0x03)
32-bit float
RW
Read
Holding
(0x03)
32-bit float
RW
Read
Holding
(0x03)
32-bit float
RW
Read
Holding
(0x03)
32-bit Integer RW
Read
Holding
(0x03)
32-bit Integer RW
Read
Holding
(0x03)
32-bit Integer RW
Read
Holding
(0x03)
32-bit float
RW
Read
Holding
(0x03)
32-bit float
RW
Read
Holding
(0x03)
Appendix D. Modbus Register Map
Modbus
Qty. of
Request
Registers
2
2
Write 0 to clear
2
When TEST MODE
enabled, force
output to value
2
2 = TEST MODE
All else = run mode
2
Measurement/Unit
Code --- 0 = off,
others see tab
2
1 = 4-20 mA,
2 = 0-20 mA,
3 = 0-2 volts
2
2
2
When TEST MODE
enabled, force
output to value
2
2 = TEST MODE,
All else = run mode
2
Measurement/Unit
Code --- 0 = off,
others see tab
2
1 = 4-20 mA,
2 = 0-20 mA,
3 = 0-2 volts
2
2
Notes
127