Commissioning
7.6 5 minute quick setup with SIWATOOL
7.6.5

Receive all data

Activate the "Receive all data" function in the communication menu.
All parameters can now been saved as a backup file on the hard disk. If a module is
replaced, the backup file can be downloaded to the new module within a few seconds. At the
time of input of the backup file, the scale is directly in the calibrated state again – without a
new calibration.
7.6.6

Checking the scale following calibration

If the scale is only used for company-internal purposes, a simple check is sufficient.
Perform the following steps:
1. The scale is unloaded and shows "0 kg".
2. Place a known reference weight on the scale.
Check the displayed value.
3. If a second known reference weight is available, place it on the scale in addition.
Check whether the scale displays the sum of the reference weights.
4. Remove the reference weights from the scale.
Check that the display is "0 kg" again.
54
SIWAREX WP231
Manual, 07/2014, A5E31238908A-02