5.2.2 Entering Calibration Data Manually (cont.)
2. Just below the probe serial number, tap the
connected.
3. To the right of the
Hygro, Pressure, Oxygen, Aux 1
(
4. For each input selected, the table below the two buttons includes data entry boxes for 2 to 16 data points. The
four buttons to the right of the table are used to edit and arrange the calibration data as required:
Sort
- Tap this button to sort the data rows in ascending order by the left column.
Edit Value
- Tap this button and then tap a text box and enter the values from the relevant Calibration Data
Sheet for each point specified, using the calculator keys on the pop-up keypad. Continue entering or
checking values for each input until you have completed each channel.
Insert Row
- Use this button along with the
Remove Row
5. After you have finished entering your probe calibration data, tap the
Configuration Menu.
moisture.IQ User's Guide
Figure 58: Probe Calibration Screen
Channel
button, tap the
Aux 2
or
).
- Use this button along with the
Channel
button to select the channel to which the probe is
Type
button to select the type of input connected to that channel
Remove Row
button to organize the rows in the table as desired.
Insert Row
button to organize the rows in the table as desired.
Chapter 5. Using the Configuration Menu
Return
button to return to the
65