Chapter 9. Specifications
9.9.3
Delta F Oxygen Cell
Type
Non-depleting coulometric electrolytic oxygen sensing cell
Available Cells
ppb O
Range
2
•
DFOX-1, 0 to 500 ppb/ 5 ppm/ 50 ppm, 1/4 VCR
+/- 3% of reading or 25 ppb
ppm O
Range
2
•
DFOX-9, 0 to 1/10/100 ppm, 1/4 VCR
•
DFOX-2, 0 to 1/10/100 ppm, 1/8 compression
+/- 3% of reading or 50 ppb
•
DFOX-3, 0 to 10/100/1000 ppm, 1/8 compression
+/- 3% of reading or 200 ppb
•
DFOX-4, 0 to 100/1000/10,000 ppm, 1/8 compression
+/- 3% of reading or 2 ppm
•
DFOX-5, 0 to 50/500/5,000 ppm, 1/8 compression
+/- 3% of reading or 1 ppm
% O
Range
2
•
DFOX-6, 0 to 5%, 1/8 compression
+/- 3% of reading or 10 ppm
•
DFOX-7, 0 to 10%, 1/8 compression
+/- 3% of reading or 20 ppm
•
DFOX-8, 0 to 25%, 1/8 compression
+/- 3% of reading or 50 ppm
Sensitivity
Less than 5 ppb (0 to 500 ppbv range)
Response Time
•
Fast response to O
•
Equilibrium time is application specific
Ambient Temperature
32°F to 120.02°F (0°C to 49°C)
104