Appendix D. Modbus Register Map - GE moisture.IQ User Manual

Moisture and oxygen analyzers
Hide thumbs Also See for moisture.IQ:
Table of Contents

Advertisement

Appendix D.
Note: The Modbus Register Map in Table 13 below has some fields in grey boxes. These are currently not supported
and will return a value of –1.0.
Function
Parameter
0 Error register,
non-latching.
Read only.
1000 Latching Error
Register
2000 Analog Output
(4-20)
Channel 1 /
Output A
Channel 1 /
Output B
moisture.IQ User's Guide
Modbus Register Map
Table 13: Modbus Register Map
Sub Parameter
ID
ModBus
Address
0
0
0
Output (percent
100
of scale)
State
102
Units
104
Type
106
Zero
108
Span
110
Output (percent
120
of scale)
State
122
Units
124
Type
126
Zero
128
Span
130
Type
Access Modbus
0 32-bit Integer
RO
1000 32-bit Integer
RW
2112 32-bit float
RW
2102 32-bit Integer
RW
2104 32-bit Integer
RW
2106 32-bit Integer
RW
2108 32-bit float
RW
2110 32-bit float
RW
2132 32-bit float
RW
2122 32-bit Integer
RW
2124 32-bit Integer
RW
2126 32-bit Integer
RW
2128 32-bit float
RW
2130 32-bit float
RW

Appendix D. Modbus Register Map

Qty. of
Notes
Request
Registers
Read Input
2
(0x04)
Read Holding
2
Write 0 to clear
(0x03)
Read Holding
2
When TEST MODE
(0x03)
enabled, force output
to value
Read Holding
2
2 = TEST MODE
(0x03)
All else = run mode
Read Holding
2
Measurement/Unit
(0x03)
Code --- 0 = off,
others see tab
Read Holding
2
1 = 4-20 mA,
(0x03)
2 = 0-20 mA,
3 = 0-2 volts
Read Holding
2
(0x03)
Read Holding
2
(0x03)
Read Holding
2
When TEST MODE
(0x03)
enabled, force output
to value
Read Holding
2
2 = TEST MODE,
(0x03)
All else = run mode
Read Holding
2
Measurement/Unit
(0x03)
Code --- 0 = off,
others see tab
Read Holding
2
1 = 4-20 mA,
(0x03)
2 = 0-20 mA,
3 = 0-2 volts
Read Holding
2
(0x03)
Read Holding
2
(0x03)
125