Installation Place - Hitachi L-2100 Instruction Manual

Table of Contents

Advertisement

2.2

Installation Place

2.2 Installation Place
When selecting an installation place for this instrument, be sure
to check the following conditions:
(1)
Operating Ambient Temperature
4 to 35 °C
Temperature variations during measurement should be
minimized. Take sufficient care to prevent dew
condensation.
(2)
Operating Ambient Humidity
45 to 85% RH
(3)
Atmosphere
(a) Well ventilated.
(b) Free from acidic or alkaline gases that could corrode
metals significantly.
(c) Free from vapors of solvents (e.g., benzene, thinner)
that could affect the paint coating on the instrument.
(4)
Other Requirements
(a) Not exposed to direct sunlight.
Exposure to direct sunlight could result in deteriorated
performance or discoloration of the instrument.
Avoid installing the instrument near a window of the
laboratory room.
(b) Not exposed to drafts directly.
(c) Free from intense vibrations and shocks.
(d) Apart from a heat source such as a gas burner,
electric heater, or oven.
(e) Apart from a strong magnetism source such as an
electronic welder, high-frequency electronic furnace, or
voltage transformer.
2 - 2