GE Amersham Imager 680 Operating Instructions Manual page 237

Table of Contents

Advertisement

If...
The instrument indicator LED lights do
not light during startup diagnosis a pow-
er cable installation failure has occurred
Operation is not stopped when
the door is opened
If...
The light source is lit or the motor is run-
ning when the door to the instrument is
open, an interlock failure has occurred
Other hardware problems
If...
The door to the instrument cannot be
opened or closed or the door cannot be
locked because a foreign object is
present in the locking section or the door
sensor
The door to the instrument cannot be
opened or closed, or the door cannot be
locked and no foreign object is present
The instrument cannot be cooled down
properly
The instrument does not appear to re-
spond to software control
Amersham Imager 680 Operating Instructions 29271308 AB
8 Troubleshooting
8.1 General problems
Then...
Connect the power cable correctly
Then...
Turn off the instrument's Power switch
immediately and contact your GE repre-
sentative
Then...
Remove the foreign object
The locking section is damaged. Contact
your GE representative
The ambient air temperature is too high.
Lower the room temperature to 28°C or
lower
The instrument is not connected to the
tablet computer
Use the flash indicator tool to identify
which instrument is connected to a cer-
tain tablet computer
237

Advertisement

Table of Contents
loading

Table of Contents