Finalize and save analysis
Step
1
2
3
4
5
6
Amersham Imager 680 Operating Instructions 29271308 AB
Function
Columns displaying:
Ch. - the detection channel
Lane - the number of the lane (change the sort order by selecting up or
down arrow in the column heading)
Band - the number of the band (from top to bottom)
Volume - the integrated signal intensity in the specific band
Lane% - the relative band volume within a lane (in percentage)
Normalized value - the band volume relative to the control band volume
Mw - the calculated Mw
Rf - the relative mobility of the band
Action
Review the summary.
If the analysis is completed, click Done.
Result: The image view appears.
To save the analysis, click Save.
Result: The image file is updated with the analysis information.
Choose a location to save the analysis by selecting the Where to save file:
drop down menu and select a location in the list.
If desired, edit the prefix of the image name by selecting the Image name
prefix: field. Type the new prefix using the appearing keypad and select the
Done button to save the changes.
If desired, add a comment by selecting the Comment: field. Type the com-
ment using the appearing keypad and select the Done button to save the
changes.
5 Operation
5.8 Analyze the image
5.8.6 Summary