3.8
Image processing adjustment
To set the EIt value
Elt value can be set in the processing parameter edit screen of the CXDI Control Software.
Enter a value in the EIt text box referring to the current EI value.
• EI (Exposure Index)
EI is an approximative indicator of the dose that reaches the detector, as calculated per IEC
62494-1 standards from the captured images.
• EIt (Target Exposure Index)
EIt is the target EI level.
• DI (Deviation Index)
DI is a logarithmic index indicating the difference between the EI and EIt.