Hitachi ChromasterUltra RS 6430 Instruction Manual page 95

Diode array detector
Table of Contents

Advertisement

Electric shock due to improper grounding!
This instrument is designed to meet the basic insulation
stipulated in the international standard IE C61010-1.
To prevent an electric shock hazard due to improper
grounding, provide a proper grounding connection.
(a)
Be sure to use the grounded 3P power cable, which is
supplied as a standard accessory for the instrument.
The use of a different type of power cable may result in
an electric shock hazard. Connect the 3P power cable
to a grounded 3P power outlet as a rule.
(b)
If a grounded 3P power outlet is not available, use a 3P
table tap with a grounding wire or a 3P-2P adapter.
In this case, be sure to provide proper grounding
connection. For grounding connection, use a screw
having a diameter of M4 or larger and a turn count of 3
or more in threading, and a wire having a thickness of
1.25 mm.
(5)
Grounding Terminal
Shall have a grounding construction for a ground
resistance of 100  or less.
APPENDIX 1 - 10
WARNING