3.1
System overview
Introduction to VIA Thaw L1000
VIA Thaw L1000 tightly controls the heating of cryopreserved samples stored in cryo-
bags until it detects that the samples are thawed. Multiple sensors constantly monitor
the condition of the sample to determine when thawing is complete.
The system operates using dry heat to minimize the risk of sample contamination.
During the thawing process, VIA Thaw L1000 records data relating to the cryobag,
monitors the thawing process, and alerts the user to the following events:
• Loading temperature limit is exceeded
• Total thaw time limit is exceeded
• End-of-thaw temperature is exceeded
• Safe heater plate temperature is exceeded
• Loss of mains power
The cryobag data records are not indicated for the support of decisions related to the
suitability of a thawed cryobag for research or manufacturing.
A complete record of the thawing process can also be uploaded and viewed on
Chronicle web platform.