Nautilus Ultrasound Transducer Top Cover Replacement - GE Corometrics 170 Series Service Manual

Hide thumbs Also See for Corometrics 170 Series:
Table of Contents

Advertisement

Serviceable Assemblies: Nautilus Ultrasound Transducer Top Cover Replacement
Nautilus Ultrasound Transducer Top Cover
Replacement
Equipment Required
7-30
In order to replace the top cover of a Nautilus Ultrasound Transducer in the field,
you will require one of two kits. Refer to
kit. Unpack the kit and ensure it contains the items listed in
something is missing, contact your GE Service Representative immediately. The
text of the instruction sheet is repeated here for informational purposes.
.
Nautilus Ultrasound Transducer
Type and Cat. No. (REF)
Loop-Style Ultrasound Transducer, 5700 KAX/LAX/MAX
Button-Style Ultrasound Transducer, 5700GAX/HAX/JAX
Transducer Case Top
Flat-Seal O-Ring
Sealing Screw, #4-40 x 5/16 in
Screw Cap, Curved
Screw Cap, Flat
Belt Knob
Screw Cap Extraction Tool
a
You will need the following equipment:
Phillips Head Torque Driver
Loctite Adhesive #454 (recommended)
Anti-Static Wristband
Static-Free Work Surface
170 Series Monitor
2000947-004
Table 7-11
and ensure you have the correct
Table 7-11. Kit Summary
Table 7-12. Packing List
Item
a
The belt knob is included in button-style
kits 2011680-001 only.
Table
7-12. If
Cable Kit
Cat. No. (REF)
2011679-001
2011680-001
Qty
1
1
6
4
1
1
Revision C