General

Revision C
Functional Checkout Procedure: General
Between Factory Service visits it is necessary that the proper operation of each
monitor be verified by performing the functional checkout procedure described in
this section. This procedure should be completed prior to initially placing the
monitor on a patient, when monitor performance needs to be verified, on a semi-
annual basis, or more frequently as dictated by your equipment maintenance and
management policies.
NOTE: Read each step of this procedure thoroughly before actually performing the
test.
Visually inspect the monitor, patient cables, and other accessories for cracks,
fissures, or other signs of wear or damage. Do not use any monitor or accessory
which appears to be worn or damaged. If unsure, contact your GE Service
Representative to arrange for evaluation, replacement, or repair of the suspect
item(s).
170 Series Monitor
2000947-004
6-3