RS-232 Connectors

RS-232 Port
Baud Rate
Mode
HP or HP with Notes
Ext. BP
8-4
Peripheral Devices: RS-232 Connectors
Two RS-232C serial interface receptacles (RJ-45) allow connecting the 170 Series
Monitor to the following devices:
Critikon Model 1846/1846SX Blood Pressure Monitor
Quantitative Sentinel/Perinatal System
Nellcor N-400 Fetal Pulse Oximeter
115-Compatible Device
Table 8-1 lists the factory default settings for each port. Refer to Customizing the
Monitor on page 4-10 for information on using the service setup mode to change
these settings.
Table 8-1. Communication Default Settings
Intended Device
1
QS System
2
Critikon Model 1846/1846SX
Select a baud rate that is compatible with the external device. The available settings:
600, 1200, 2400, 4800, 9600, 19,200, and 384,000 bps.
Select a communications mode compatible with the external device.
Select HP or HP w/notes for connecting to a Marquette Quantitative/Perinatal
System or another central system that uses the Hewlett Packard communications
protocol. HP mode does not print central station annotations on the fetal monitor
strip chart, whereas HP w/notes does.
Select ext. BP for connecting to a Critikon Model 1846/1846SX Blood Pressure
Monitor.
170 Series Monitor
2000947-004
Mode
Baud Rate
HP
1200
ext. BP
600
Revision C