Connecting The Simulator - GE Corometrics 170 Series Service Manual

Hide thumbs Also See for Corometrics 170 Series:
Table of Contents

Advertisement

Connecting the Simulator

6-6
Functional Checkout Procedure: Connecting the Simulator
This part of the procedure prepares the simulator for use.
NOTE: You must use a Model 325 Simulator for the functional checkout
procedure. (Monitors in the 170 Series do not work with Model 305
Simulators.)
1.
Ensure the Model 325
2.
Connect the Model 325 Simulator's power cord to the power receptacle on the
rear panel of the simulator; plug the other end of the power cord into a properly
grounded wall outlet of appropriate voltage.
3.
Ensure the 170 Series Monitor is turned off.
4.
Connect the simulator interconnect cable's 50-pin end to the simulator's
connector.
Monitor
5.
Connect the sub-cables of the other end of the simulator interconnect cable into
the color-coded connectors on the monitor/adapter:
the procedure).
6.
Turn on the Model 325 Simulator. Verify that the green
illuminates.
7.
Turn on the 170 Series Monitor.
170 Series Monitor
2000947-004
switch is in the off position.
Power
Fetal
and
(as required by
US
UA
indicator
Power
Revision C