GE Moisture Image 1 Series Service Manual page 30

Hygrometer
Hide thumbs Also See for Moisture Image 1 Series:
Table of Contents

Advertisement

Performing an MH
Calibration Test/
Adjustment (cont.)
Installing Optional Features
Calibration Procedure (cont.)
3. Reprogram the meter with the new (corrected) low reference value
(if required) as described in Entering High and Low Reference
Values in Chapter 2 of the Programming Manual.
4. Verify that the probe cable is not connected to the probe.
5. Note the zero reference readings and verify that the readings are
now within ±0.0003 MH.
6. Fill out a new high and low reference sticker with the final low
reference value and/or record the information on the Program
Information List provided in Appendix A of the Startup Guide.
Make sure you record the information below:
HIGH REF = ORIGINAL VALUE
LOW REF = NEW CORRECTED VALUE
ZERO REF = ORIGINAL RECORDED VALUE
7. Reconnect the probe to the cable.
Note: If cables are changed in any way, repeat this procedure for
maximum accuracy.
The Series 1 is now ready for operation. Proceed to Chapter 2, Basic
Programming, of the Programming Manual for instructions.