GE Moisture Image 1 Series Service Manual page 47

Hygrometer
Hide thumbs Also See for Moisture Image 1 Series:
Table of Contents

Advertisement

Checking and
Replenishing the
Electrolyte in the
Delta F Oxygen Cell
(cont.)
Troubleshooting and Maintenance
To replenish the electrolyte:
Once the oxygen cell receives the initial charge of electrolyte, you
should monitor the level regularly. DO NOT let the fluid level drop
below the "MIN" level mark on the window.
THE ELECTROLYTE CONTAINS A STRONG
CAUSTIC INGREDIENT AND CAN BE HARMFUL IF
IT COMES IN CONTACT WITH THE SKIN OR THE
EYES. FOLLOW PROPER PROCEDURES FOR
HANDLING THE CAUSTIC (POTASSIUM
HYDROXIDE) SOLUTION. CONSULT YOUR
COMPANY SAFETY PERSONNEL.
To raise the fluid level in the reservoir, add DISTILLED WATER
slowly in small amounts. Check the level as you add the distilled
water, making sure you do not overfill the reservoir. The electrolyte
mixture should cover approximately 60% of the min/max window.
!WARNING!
June 2003
2-15