4.6 Error Messages

Message
ROM ERROR
PRESS "CL" KEY TO
CLEAR MESSAGE.
RAM ERROR
PRESS "CL" KEY TO
CLEAR MESSAGE.
PARAMETER ERROR
PARAMETERS ARE
SET TO DEFAULT
VALUE.
LOG INFORMATION
ERROR
LOG INFORMATION IS
INITIALIZED.
EEPROM ERROR
INSTRUMENT INFO. IS
INITIALIZED.
Ex (Em) SIDE OF WL
DRIVE MECHANISM
ERROR
Ex (Em) SIDE
ACCURACY (254 nm)
NOT DETECTED
LAMP OFF BY
OVERHEAT!
PRESS "CL" KEY TO
CLEAR MESSAGE.
LAMP OFF BY
EXTERNAL INPUT
PRESS "CL" KEY TO
TURN ON.
Description
ROM is faulty.
RAM is faulty.
Analytical parameters are
inadequate.
Log information is
inadequate.
Instrument parameters are
inadequate.
Failure in detecting Ex or Em
wavelength origin during self-
diagnosis test (initialization).
Wavelength error exceeds
10 nm during wavelength
accuracy check.
Abnormal temperature
detected
Xe lamp turned off to prevent
its overheating.
Lamp has been turned off
due to an external contact
input. Busy status is
assumed.
4 - 33
Remedy
1) Press CL key, and 1st monitor
screen returns.
2) Contact your nearest service office
of Hitachi High-Tech Science sales
representative.
1) Press CL key, and 1st monitor
screen returns.
2) Contact your nearest service office
of Hitachi High-Tech Science sales
representative.
1) Press CL key, and 1st monitor
screen returns.
2) Turn on power again.
3) Contact your nearest service office
of Hitachi High-Tech Science sales
representative.
1) Press CL key, and 1st monitor
screen returns.
2) Turn on power again.
3) Contact your nearest service office
of Hitachi High-Tech Science sales
representative.
1) Press CL key, and 1st monitor
screen returns.
2) Turn on power again.
3) Contact your nearest service office
of Hitachi High-Tech Science sales
representative.
1) Cannot be released by key
operation.
2) Turn on power again.
3) Contact your nearest service office
of Hitachi High-Tech Science sales
representative.
1) Carry out troubleshooting (4.5).
2) Contact your nearest service office
of Hitachi High-Tech Science sales
representative.
1) Press CL key, and 1st monitor
screen returns.
2) Contact your nearest service office
of Hitachi High-Tech Science sales
representative.
1) Press CL key, and 1st monitor
screen returns.
2) Use UTLTY key to light up the lamp
again.
1) Release the key lock if it has been
activated.
2) Press CL key and light up the lamp
again.