Change The Settings - Panasonic MLS-3751 Operating Instructions Manual

High-pressure steam sterilizer
Table of Contents

Advertisement

How to Use

8. Change the settings

It is possible to change the settings (sterilization temperature, sterilization
time, melting temperature, melting time, keep warm temperature, exhaust
temperature and exhaust rate) according to need.
The settings that have been changed are saved in memory even when the
power switch is turned OFF. (Once the operation is started, the settings of
the sterilization temperature and melting temperature can not be changed.
The sterilization time, etc. can be changed but not saved in memory.)
(Changeable range: See page 18.)
1) Press the
button.
Setting menus flicker.
The setting menu moves each time this button is pressed.
2) Press the
or
MEMO
The setting value is incremented or decremented each time the
or
button is pressed.
If the button is kept pressed, the value is rapidly incremented or
decremented and it stops finally at the maximum or minimum value.
3) Press the
button to flicker the next menu.
Press the
or
Press the
button.
Repeat this procedure.
Press the
or
For setting menu to be left unchanged, simply press the
4) Finally, press the
When the setting menu is entered, a buzzer sound is heard and
the changed setting is saved in the program.
The flickering status is left standing for 1min., the changed setting
is not updated and the product turns to be standby status.
(Standby status : See page 16.)
Each time the button is pressed, the setting menus flickers one by one in the sequence as follows:
1. Liquid Sterilization
Cycle
Sterilization temperature
Sterilization time
Exhaust temperature
Exhaust rate
20
Beep
button to change the setting.
button to change the setting.
button to change the setting.
button.
2. Sterilization/Keep
warm Cycle
Sterilization temperature
Sterilization time
Exhaust temperature
Exhaust rate
Keep warm temperature
Beep
button.
3. Melting/Keep
warm Cycle
Melting temperature
Melting time
Exhaust rate
Keep warm temperature
( Continued )
4. Instrument
Sterilization Cycle
Sterilization temperature
Sterilization time
Exhaust rate