Set The Power Switch To On; Open The Cover - Panasonic MLS-3751 Operating Instructions Manual

High-pressure steam sterilizer
Table of Contents

Advertisement

2. Set the power switch to ON

The digital display of the control panel lights up. (The boiling point
setting temperature appears on the digital display
appears on the digital display
Digital display
Cover being CLOSE
Settings
set temperature and
(
sterilization time/cycle) are indicated.
Digital display : Sterilization temperature
Digital display : Sterilization time and
cycle indicated alternately
This status is referred to as Standby
Status
.

3. Open the cover.

1) Be sure to check the following points before opening
the cover.
If you attempt to open the cover by force, trouble may
be caused with the product.
Power switch is turned ON.
Pressure gauge indicates 0 MPa.
COVER LOCK lamp goes out.
2) While pressing down the handle,
3) Pull the movable handle toward you.
The cover can not be unlocked unless the handle
is pressed down.
4) Open the cover.
for the first 2 seconds.)
To check the temperature inside
the chamber,Press the
and
The current temperature is indicated
while these buttons are kept pressed.
Digital display : Temperature inside
the chamber
Digital display : Temperature of
heating water
If the temperature is lower than 25 °C,
appears.
(The cover can not be opened unless the power switch is turned ON)
and - - -
buttons simultaneously.
The figure shown above is for the
case when the boiling point set
temperature is 100 °C.
Cover being OPEN
The current temperature of the
heating water in the chamber is
indicated.
The figure shown above is for
the case when the temperature
of the heating water in the
chamber is 30 °C.
Make sure that the
COVER LOCK lamp goes out.
Make sure that the pressure
gauge indicates O .
Make sure that the power
switch is turned ON.
Handle
Movable
2)
handle
4)
3)
17