Erase
Note: Erase will delete ALL calibration and user settings. The
HygroTrace will be incapable of performing accurate
measurements until this data is restored.
Note: This function is provided to assist GE Sensing service
personnel in maintaining the product.
Heater Test
To turn off the test, select OFF and press Enter. The temperature
in the upper right corner of the screen returns to its original value.
To resume normal operation, select RESUME and press Enter.
Press Cancel to return to the Service Menu.
32
To erase the EPROM, select Erase,
press Enter and the following
message appears at the top of the
screen for a few seconds.
+
Select YES or NO and press Enter, or
press Cancel. If you select NO or
press Cancel, the unit returns to the
previous screen. If you select YES,
the following message appears.
+
When the erasure is complete,
Param Storage reappears. To return
to the Service Menu, press Cancel.
To perform a test of the heater, select
Heater Test..., press Enter and a
screen similar to the following
appears:
+
Use the arrow keys to select the
appropriate temperature to test, and
press Enter. The temperature in the
upper right corner of the screen
should approximate the one selected.
+