DEFIBRILLATION ELECTRODES (PADS)

On the Responder AED PRO, an audible and visual alert will indicate after the self-test if the pads are missing, unplugged
or damaged.
CAUTION: Possible Improper AED Performance
Using pads that are damaged or expired may result in improper AED performance.
PAD INSTALLATION
CAUTION: Use only Approved Equipment
Using batteries, pads, cables, or optional equipment other than those approved by
General Electric may cause the AED to function improperly during a rescue.
CAUTION: Possible Improper AED Performance
Using pads that are damaged or expired may result in improper AED performance.
DIRECTIONS FOR USE:
1.
Do NOT open until ready to use, short term use only.
2.
Ensure the skin site is clean and dry.
3.
Separate one pad from liner.
4.
Place one pad on skin in either location.
5.
Peel and place remaining pad.
2023491-201 Rev B
The defibrillation pads come in a ready-to-use, sealed package containing one pair of
self-adhesive pads with an attached cable and connector. The pads are disposable
and should be discarded after each rescue. The pads have a limited shelf life and
shall not be used beyond the expiration date. Keep a fresh, unopened pair of pads
plugged into the AED at all times. Refer to the pad package label for operation
temperatures.
1.
Open the lid of the AED.
2.
Place the package into the lid so that the expiration label is visible
through the clear window on the lid. The expiration date of the pads will
then be readable without opening the lid of the AED.
3.
Match the color of the connectors (red to red), slightly lift the tab of the
pad socket and then plug the pad connector into the AED case as shown
in the drawing.
4.
Tuck the excess cable length in the bottom holder as shown in the
drawing. With the package completely secured to the AED lid, close the
lid.
5.
Make sure the expiration date is visible through the clear window of the
lid.
Make sure that the STATUS INDICATOR is GREEN.