SVT RATE
All rhythms with rates between the Detection Rate and SVT Rate will be screened through a number of SVT
Discriminators to classify them into VF/VT or SVT. Rhythms classified as SVT between the two set rates are not
shockable. All SVT rhythms above the rates will be classified as shockable. The SVT Rate must be greater than the
Detection Rate and is selectable by Service between 160 and 300 bpm or, "NO THERAPY FOR SVT" can be selected by
Service using the ServiceLink software.
RESCUE PROTOCOL
The AED rescue protocol is consistent with the guidelines recommended by the American Heart
Association (AHA)
on Resuscitation (ILCOR).
Upon detecting a shockable cardiac rhythm, the AED advises the operator to press the SHOCK button to deliver a shock
and then advises the operator to start CPR.
Note: The standard CPR protocol of 120 seconds can be modified from 60 to 180 seconds in ServiceLink

OPERATOR TRAINING REQUIREMENTS

Persons authorized to operate the AED must have all of the following minimum training.
Defibrillation training and other training as required by state, province, or country regulations.
Training on operation and use of the AED.
Additional training as required by the physician or Medical Director.
A thorough understanding of the procedures in this manual.
Note: Keep valid certificates of training and certification as required by state, province, or
country regulations.
1
"Guidelines 2005 for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care" American Heart Association; Circulation
Vol112,Issue 24 Suppl. Dec 13, 2005
2023491-201 Rev B
1
European Resuscitation Council (ERC) and the International Liaison Committee
© 2007 General Electric Company
SECTION 2 - INTRODUCTION
Page 13 of 33

Advertisement

Table of Contents
loading

This manual is also suitable for:

2023681

Table of Contents