GE AED Pro Service Manual page 32

Hide thumbs Also See for AED Pro:
Table of Contents

Advertisement

Table A2 – Low Variable Energy Waveform Responder AED Pro (all values
are typical
Patient's
Impedance
(Ohms)
25
50
75
100
125
Table A3 – High Variable Energy Waveform Responder AED Pro (all values
are typical
Patient's
Impedance
(Ohms)
25
50
75
100
125
ENERGY LEVELS AND PATIENT IMPEDIANCE
The Biphasic Truncated Exponential (BTE) waveform utilizes variable energy. The actual energy delivered will vary with
the patient's impedance and the device will deliver a shock when impedance is between 25-180 Ohms. Energy will be
delivered at three different levels referred to as ultra-low variable energy, low variable energy, and high variable energy as
shown in the above waveform tables.
2023491-201 Rev B
Phase 1
Voltage
Duration
(Volts)
(ms)
1570
3.3
1600
4.5
1610
5.8
1615
7.0
1620
8.3
Phase 1
Voltage
Duration
(Volts)
(ms)
1885
3.3
1920
4.5
1930
5.8
1940
7.0
1945
8.3
© 2007 General Electric Company
SECTION 4 – TECHNICAL DATA
Phase 2
Voltage
Duration
(Volts)
(ms)
825
3.2
1030
3.2
1105
3.2
1150
3.2
1170
3.2
Phase 2
Voltage
Duration
(Volts)
(ms)
990
3.2
1240
3.2
1325
3.2
1380
3.2
1405
3.2
Energy
(Joules)
200-250
170-210
120-160
150-180
140-170
Energy
(Joules)
265-360
235-320
215-295
200-270
190-260
Page 31 of 33