Internal Ua Troubleshooting - GE Corometrics 170 Series Operator's Manual

Fetal monitor
Hide thumbs Also See for Corometrics 170 Series:
Table of Contents

Advertisement

Internal UA Troubleshooting

Problem
Internal pressure not measuring correctly.
9-6
Troubleshooting: Internal UA Troubleshooting
Table 9-5. Internal UA Troubleshooting
Probable Cause
Transducer not properly connected to
monitor.
Air bubble in dome; or catheter blocked.
Dome is cracked.
Strain gauge not at same height as
catheter tip.
Catheter has fallen out of place.
Catheter or strain gauge not zeroed.
Service required.
Blockage in fluid-filled catheter.
Fetus pressing directly on catheter.
Defective strain gauge or catheter.
Service required.
170 Series Monitor
2003023-001
Possible Solution
Ensure transducer is securely attached to
monitor.
Flush dome and catheter.
Replace dome.
Adjust strain gauge height.
Replace catheter.
Calibrate catheter or strain gauge.
Call for service.
Flush catheter. Re-zero. Replace catheter
if necessary.
Reposition by twisting catheter.
Replace strain gauge or catheter.
Call for service.
Revision D

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents