1.6 Alarms Option Card
Alarm relays may be added to the Model DF868 by installing an Alarms Option Card in one (or more) of the six
expansion slots. Each option card includes three alarm relays, which are designated as A, B, and C. To test the alarm
relays, connect an ohmmeter to the option card terminal block as shown in Figure 4 below. Refer to the menu map in
Figure 6 on page 13 as a guide.
For this discussion, assume that the option card has been installed in
1.6.1 Accessing the Calibration Menu
[CAL]
1. Press the
key to enter the Calibration Program.
[Fx]
Slot x
2. Press
to select
[F1]-[F3]
3. Press
to select alarm relay A, B or C, respectively.
DigitalFlow™ Multipurpose Ultrasonic Liquid Flowmeter Service Manual
9
C
NC
C
COM
C
NO
B
NC
B
COM
B
NO
A
NC
A
COM
A
NO
1
Figure 4: Typical Ohmmeter Connections
. (The option bar will include a slot listing for each installed option card.)
Terminal Block (Option Card)
–
Ohmmeter
+
Slot x
.
Chapter 1. Calibration
9