Fuse Replacement

Fuse Replacement
Equipment required:
• Small flathead screwdriver
Warning
Fire Hazard For continued protection against fire hazard, replace fuses only with fuses of the
same type and rating as the original fuse.
1.
Turn off the system. Disconnect the power cord from the wall receptacle and the rear
panel of the system for easier access to the adjacent fuse drawer.
2.
To release the fuse drawer, insert a small flathead screwdriver into the slot on the
drawer above the power cord receptacle, then slide the drawer out.
3.
Replace each fuse with one of the same type and rating.
4.
Slide the fuse drawer into its slot until it snaps into place.
5.
Connect the power cord to the rear panel.