Operation and Service Manual
Kendall SCD
700 Sequential Compression System
700 Système de compression séquentielle
- Manuel d'utilisation et d'entretien
700 Sequenzielles Kompressionssystem
- Bedienungs- und Servicehandbuch
Sistema di compressione sequenziale 700
- Manuale d'uso e di manutenzione
Sistema de compresión secuencial 700
- Manual de funcionamiento y mantenimiento
700 Sekventiellt kompressionssystem
- Användar- och servicehandbok
700 sequentieel compressiesysteem
- Bedienings- en onderhoudshandleiding
Sistema de Compressão Sequencial 700
- Manual de Funcionamento e Assistência
700 jaksoittainen kompressiojärjestelmä
- Käyttö- ja huolto-ohjekirja
700 Sekventielt kompressionssystem
- Bruger- og servicevejledning
Σύστημα διαδοχικής συμπίεσης 700
- Εγχειρίδιο λειτουργίας και σέρβις
Sekvenční kompresní systém 700
- Uživatelská a servisní příručka
700 Szekvenciális kompressziós rendszer
- Kezelési és szervizelési kézikönyv
700 Система терапевтическая для
последовательной компрессии
- Руководство по эксплуатации и
обслуживанию
System stopniowanego ucisku 700
- Podręcznik obsługi i serwisu
700 Sıralı Kompresyon Sistemi
- Çalıştırma ve Servis El Kitabı
700 Sekvensielt kompresjonssystem
- Bruker- og servicehåndbok
Sekvenčný kompresný systém 700
- Príručka na obsluhu a servis
Sistem de compresie secvenţială 700
- Manual de operare şi întreţinere
Система за последователна компресия 700
- Ръководство за работа и сервиз

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Kendall SCD 700 Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Covidien Kendall SCD 700 Series

  • Page 1 Operation and Service Manual Kendall SCD ™ 700 Sequential Compression System 700 Système de compression séquentielle - Manuel d’utilisation et d’entretien 700 Sequenzielles Kompressionssystem - Bedienungs- und Servicehandbuch Sistema di compressione sequenziale 700 - Manuale d’uso e di manutenzione Sistema de compresión secuencial 700 - Manual de funcionamiento y mantenimiento 700 Sekventiellt kompressionssystem - Användar- och servicehandbok...
  • Page 2: Table Of Contents

    Section VI - Test Methods and Calibration ................EN-19 Test Mode Look up Chart............................EN-20 Test Mode T1 - Burn-In ............................EN-20 Test Mode T2 - General Function Test ......................... EN-20 Test Mode T3 - Pressure Transducer Calibration ....................EN-20 Kendall SCD 700 Series...
  • Page 3 Adjustable Bed Hook (Removal/Installation) ..................... EN-24 Section VIII - Parts Listing ....................EN-25 Section IX - Specifications ....................EN-26 Kendall SCD 700 Series Compression System ..................... EN-26 Section X - Schematics ....................EN-29 Figure 9 - Parts Assembly Diagram – Exploded view (Page 1 of 2) ..............EN-29 Figure 9 - Parts Assembly Diagram (front enclosure)–...
  • Page 4: Indications

    5. Suspected pre-existing deep venous thrombosis. Foot Compression The Kendall SCD 700 series compression system may not be recommended for use with foot cuffs on patients with the following: 1. Conditions where an increase of fluid to the heart may be detrimental.
  • Page 5: Cautions And Warnings

    55°C Manufacturer Store between these temperatures. -4°F -20°C WEEE (Waste from electrical Protection against fluid ingress: spraying water and electronic equipment) Protection against fluid ingress: spraying water Protective earth (ground) and particulates Equipotential ground point Kendall SCD 700 Series EN-2...
  • Page 6: Front Panel Display

    Constructed from recyclable materials. Device contains phthalates. Front Panel Display Item Explanation Item Explanation Power On Indicator Battery Status Indicators 1-3 Power On/Standby Button Right Arrow Button A - Button B - Button AC Power/Battery Charging Indicator EN-3 Kendall SCD 700 Series...
  • Page 7: Section I - General Operating Instructions

    Also after startup, the controller immediately begins conducting the Garment Selection and Verification procedure at each port to determine if the garments have been properly attached to the controller: Kendall SCD 700 Series EN-4...
  • Page 8 In this case, either the system can be turned off and restarted or the corresponding A and B Button(s) can be pressed to confirm problem resolution and operation will continue without having to power the controller down and restarting. EN-5 Kendall SCD 700 Series...
  • Page 9: Normal Operation And Pressure Adjustment

    • The time between compressions on the same limb will never be shorter than twenty seconds or longer than sixty seconds. • If both controller ports are being used, then the longer of the two measurements will be used to adjust the time between cycles. FIGURE 3 Shutdown To terminate the operation, press the Power On/Standby button on the device. Kendall SCD 700 Series EN-6...
  • Page 10: Garment Compatibility

    Garment Compatibility The Kendall SCD 700 series compression system is designed for use with Kendall SCD garment Reorder Codes: Kendall SCD Sequential Compression Express Sleeves Comfort Sleeves 9529 Knee Length Medium 74010 Thigh Length X-Small 9530 Thigh Length Medium 74011 Thigh Length Small...
  • Page 11: Section Ii - Compliance Meter

    Introduction The Kendall SCD 700 series controller has a feature called the Compliance Meter, that can be used to monitor the amount of time a patient receives compression therapy either by shift, day or during an entire hospital stay. This unique feature operates in the background, so that it does not interrupt daily normal operation.
  • Page 12: Accessing The Compliance Meter

    • Pressing the A button (shift selector) allows the user to select a time duration of interest. Note that the selection indicator moves with each button press. • To determine the amount of therapy a patient has received over the most recent 8 hours, for example, select the ‘8’ on the shift selector. • To determine the amount of therapy a patient received over the most recent 24 hours, for example, select the ‘ 24’ on the shift selector. • Be aware, if the amount of elapsed time has not yet reached the time selected on the shift selector, then the actual elapsed time will display in the bottom number. • Note that after 30 seconds of inactivity, the Compliance Meter will return to the normal therapy screen. EN-9 Kendall SCD 700 Series...
  • Page 13 Kendall SCD 700 Series EN-10...
  • Page 14: Section Iii - Battery Operation

    Section III - Battery Operation The Kendall SCD 700 series compression system is designed to operate normally on AC line power or DC battery power without interruption. There are three Battery Status Indicator LED’s used to represent the charge level of the battery.
  • Page 15: Unit Powered Off (Charging When Plugged In

    50% and kept near 25°C (77°F) if prolonged storage is necessary. Battery Warnings The Kendall SCD 700 series compression system battery pack contains Lithium Ion (Li-Ion) battery cells and must be used properly for safety and to maintain optimal performance. • Store spare battery packs between –20°C (-4°F) and 60°C (140°F).
  • Page 16: Section Iv - Fault Conditions And Troubleshooting

    A yellow triangle indicates a low concern error. If the triangle is red it is indicative of an error related to a pressure that is high in an abnormal way. If the error persists, then the controller requires service. Kendall SCD 700 Series EN-13...
  • Page 17 Low Pressure (Foot Cuffs) User Resettable Foot Cuff pressure is less Check for leaks in the cuff than 125 mmHg after 10 or the tube connections. consecutive cycles. EN-14 Kendall SCD 700 Series...
  • Page 18 The pressure in a garment Check tubing for kink or is greater than 20 mmHg occlusion. Check garment at the end of any vent application (too loose or period. tight). Service Technician only: Check for kinked internal tubing. Kendall SCD 700 Series EN-15...
  • Page 19: Section V - Service And Maintenance

    It is not to be construed as authorization to perform warranty repairs. Unauthorized service will void the warranty. Introduction The Kendall SCD 700 series controller contains no user serviceable parts. User maintenance is covered in the sections that follow. All other maintenance must be performed by technically qualified service personnel. EN-16...
  • Page 20: Warranty And Factory Service

    Service technicians should be familiar with the operator’s portion of this manual and the operating principles of the Kendall SCD 700 series compression system. If a controller is to be returned to Covidien for service, a description of the operating conditions and the fault code displayed should accompany the unit. The fault codes displayed by the controller are useful in diagnosing service problems.
  • Page 21: Fuses

    General Function Test (Test Mode T2) The expected service life of the Kendall SCD 700 series controller is 5 years. However, the life of the controller can be extended indefinitely by replacing components if they fail. Refer to the spare parts listing within this Operation and Service Manual.
  • Page 22: Tube Set Cleaning

    PC Board. Covidien does not recommend any attempt to repair printed circuit boards. In manufacturing, extensive testing is performed that cannot be duplicated in the field without specialized equipment. Improper repair could result in patient or user hazards.
  • Page 23: Test Mode Look Up Chart

    Test Mode T3 - Pressure Transducer Calibration Note: The transducer used in the Kendall SCD 700 series compression system is a state-of -the-art, highly precise and virtually drift free device. Factory calibration certification is void if the case is opened. Recalibration is rarely required and should be done only when necessary.
  • Page 24: Test Mode T4 - Pressure Transducer Calibration Verification

    Test Mode T4 - Pressure Transducer Calibration Verification Note: The transducer used in the Kendall SCD 700 series compression system is a state-of-the-art, highly precise and virtually drift free device. Factory calibration certification is void if the case is opened. Recalibration is rarely required and should be done only when necessary.
  • Page 25: Test Mode T7 - Manufacturing Test

    Battery Pack (Removal / Installation - see Figure 9) • Disconnect the battery wiring harness from the main CPU board, cut wire ties as required, noting their locations for reassembly. • Slide the battery pack out of its pocket. • Installation is the reverse of removal. Compressor (Removal / Installation - see Figure 11) • The compressor is not a user serviceable component. Do not disassemble. Do not oil. The compressor is held in place by friction of its molded foam enclosure. • Disconnect the compressor wiring harness from the controller Board on the front case and cut wire ties as required, noting their locations for reassembly. EN-22 Kendall SCD 700 Series...
  • Page 26: Muffler (Removal / Installation

    • To remove the fan, disconnect the 2-pin fan connector from the power board. Cut wire ties as required, noting their locations for reassembly. • Remove the three screws from the fan and remove it from the enclosure. • Installation is the reverse of removal. Use care to ensure that the direction of flow is correct. The fan is intended to pull air through the power cord door. Note the molded arrow in the fan case showing the flow direction. • For optimum cooling and quietness, use only Covidien replacement fans. Kendall SCD 700 Series EN-23...
  • Page 27: Main Cpu Board And Graphical Display (Removal/Installation - See Figure 9

    • Place the controller on its front on a non-marring surface. • Grasp both the left and right sides of the bed hook at the pivot point. Pull the bed hook out while simultaneously rotating the bed hook up toward the top of the controller. • Torsion springs may snap free or slide off the mandrel of the pivot. Use care so the torsion springs do not release hazardously. Note their location for ease of reassembly. • When reinstalling reverse these steps, being careful to start reinsertion with the bed hook rotated upward toward the top of the controller. EN-24 Kendall SCD 700 Series...
  • Page 28: Section Viii - Parts Listing

    Exhaust Filter 1036056 LCD Display* 1029099 Main CPU Printed Circuit Assembly 1056673 LCD Display* 1058683 Main CPU Printed Circuit Assembly 1062408 *When ordering an LCD Display, ensure compatibility with the main CPU printed circuit board. Kendall SCD 700 Series EN-25...
  • Page 29: Section Ix - Specifications

    Section IX - Specifications Kendall SCD 700 Series Compression System Safety Standards Built to UL60601-1, CSA-C22.2 No. 601.1-M90, CSA C22.2 N0. 60601-1: 2008, JIS T 0601-2-204, EN60601-1, ANSI/AAMI ES60601-1:2005 and IEC 60601-1-2:2007 Standards UL Classified File # E189131 and E351453 Device Classification Class I - Equipment Internally Powered, Portable Type BF Applied Parts Not AP...
  • Page 30 Guidance and manufacturer’s declaration – electromagnetic immunity The Kendall SCD 700 series is intended for use in the electromagnet environment specified below. The customer or the end user of the Kendall SCD 700 series should assure that it is used in such an environment.
  • Page 31 Guidance and manufacturer’s declaration - electromagnetic emissions The Kendall SCD 700 series controller is intended for use in the electromagnetic environment specified below. The customer or the user of the Kendall SCD 700 series should assure that it is used in such an environment.
  • Page 32: Section X - Schematics

    Section X - Schematics Figure 9 - Parts Assembly Diagram – Exploded view (Page 1 of 2) EN-29 Kendall SCD 700 Series...
  • Page 33: Figure 9 - Parts Assembly Diagram (Front Enclosure)– Exploded View (Page 2 Of 2)

    Figure 9 - Parts Assembly Diagram (front enclosure)– Exploded view (Page 2 of 2) Kendall SCD 700 Series EN-30...
  • Page 34: Figure 10 - Pneumatic & Electrical Schematic

    Figure 10 - Pneumatic & Electrical Schematic EN-31 Kendall SCD 700 Series...
  • Page 35: Figure 11 - Rear Enclosure View

    Figure 11 - Rear Enclosure View Equipotential Lug Location Muffler Power Compressor Battery Supply Assembly Kendall SCD 700 Series EN-32...
  • Page 36: Figure 12 - Front Enclosure View

    Figure 12 - Front Enclosure View Main CPU Board EN-33 Kendall SCD 700 Series...
  • Page 37 Данни за вещество, което не се съдържа или не се намира в продукта или опаковката. IFU No. 1056225 COVIDIEN, COVIDIEN with logo, and Covidien logo are U.S. and internationally registered trademarks of Covidien AG. ™*Trademark of its respective owner. Other brands are trademarks of a Covidien company.
  • Page 38 NUMBER OF COLORS OPTION A BLACK PMS 281C PMS 292C XXXX XXXX XXXX Kendall SCD Product Brand: 15 Hampshire Street, Mansfield, MA 02048 700 Sequential Compression System Product Descriptor: 29525B XXXX XXXX XXXX XXXX XXXX XXXX REF #: TINTS GBU-SPECIFIC INFORMATION Vascular Therapies/Compression GBU &...