GE LOGIC C2 Series Technical Publication page 65

Ultrasound system
Table of Contents

Advertisement

Electrical Requirements for Docking Carts
Table 3-10:
Docking Cart for LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid| e/Vivid iq Power
PARAMETER
Voltage Range (Docking
Cart used for LOGIQ e/
LOGIQ e Vet R7.x.x and
below, LOGIQ i and Vivid
e)
Voltage Range (Docking
Cart used for LOGIQ e
R8.x.x or higher)
Vivid iq Docking Cart
Power
Line Frequency
Power Transients
Decaying Oscillation
Ultrasound System
Direction 5444964-100 English Rev. 5
General Ultrasound system requirements
Requirements
AREA
100-120V~
220-240V~
100-240V~
100-240V~
All applications
All applications
All applications
All applications
Common Service Information
LIMITS
500VA
500VA
350 VA
190VA ~ 350 VA
More than or equal to 750
VA
50/60Hz (±2Hz)
Less than 25% of
nominal peak voltage for
less than 1 millisecond
for any type of transient,
including line frequency,
synchronous,
asynchronous, or
aperiodic transients.
Less than 15% of peak
voltage for less than 1
millisecond.
FREQUENCY
50/60 Hz
3-13