GE LOGIC C2 Series Technical Publication page 64

Ultrasound system
Table of Contents

Advertisement

Site Preparations
Electrical requirements for Laptop Ultrasound Systems (continued)
Table 3-8:
Product
AC/DC Converter for
Vivid i/Vivid iN and
Vivid q/ Vivid qN
Adapter for Vivid e
Table 3-9:
Product
Voluson i/e
Voluson Station /
Voluson Dock Cart
3-12
Electrical Requirements - Vivid Laptops
Voltage
Tolerance
100 - 240 VAC
±10%
100 - 240 VAC
±10%
Electrical Requirements - Voluson Laptops
Voltage
Tolerance
100 - 240 VAC
±10%
100-130 VAC
220-240 VAC
Ultrasound System
Current or
Power
Consumption
0.5 to 1.0 A
130 VA max. 1.3 A
(max.)
Current or
Power
Consumption
120 VA
320 VA
Common Service Information
Direction 5444964-100 English Rev. 5
Frequency
50-60 Hz
50/60 Hz
Frequency
50, 60 Hz
(±2%)