GE LOGIC C2 Series Technical Publication page 62

Ultrasound system
Table of Contents

Advertisement

Site Preparations
Electrical requirements for Console Ultrasound Systems (continued)
Table 3-6:
Product
Voluson® 730
Voluson E6 /
Voluson E8
model-type <=BT13
Voluson E6 /
Voluson E8
model-type >BT13
Voluson E10,
>BT15
Voluson P6 /
Voluson P8
Voluson S6 /
Voluson S8
Voluson S10
3-10
Electrical Requirements - Voluson Consoles
Voltage
Tolerance
100 VAC
110 VAC
115 VAC
130 VAC
±10%
230 VAC
240 VAC
100 - 130 VAC
220 - 240 VAC
100-240 VAC
100-240 VAC
100-130 VAC
200-240 VAC
100-120 VAC
220-240 VAC
Ultrasound System
Current or
Power
Consumption
10.1 A
9.2 A
8.8 A
7.8 A
50, 60 Hz (±2Hz)
4.4 A
4.2 A
10.3 - 7.7 A
4.5 - 4.2 A
800 VA
900 VA
900 VA
900 VA
Common Service Information
Direction 5444964-100 English Rev. 5
Frequency