Environmental Conditions - Sanyo MCO-18AC UV Instruction Manual

Co2 incubator
Table of Contents

Advertisement

ENVIRONMENTAL CONDITIONS

This equipment is designed to be safe at least under the following conditions (based on the IEC
61010-1):
■ Indoor use;
■ Altitude up to 2000 m;
o
o
■ Ambient temperature 5
C to 40
C
o
■ Maximum relative humidity 80% for temperature up to 31
C decreasing linearly to 50% relative humidity
o
at 40
C;
■ Mains supply voltage fluctuations not to exceed ±10% of the nominal voltage;
■ Other supply voltage fluctuations as stated by the manufacturer;
■ Transient overvoltages according to Installation Categories (Overvoltage Categories) II; For mains
supply the minimum and normal category is II;
■ Pollution degree 2 in accordance with IEC 60664.