SETTING OF KEY LOCK (MENU/Std-by)
Caution: The buzzer rings for a long time when a wrong password is input. Input a correct password.
The password for key unlock must be shared and administered by all users on this product.
The setting of key unlock when shipped from the factory is "0000".
Refer to page 40 for changing the password.

PROCEDURES OF SETTING VALUE

Setting of chamber temperature

Procedures of changing chamber temperature from 25℃ to -10℃ are shown below. As for other values,
the same procedure is applied.
Procedures of changing chamber temperature from 25℃ to -10℃
1
2
Press the menu button (MENU).
Select Std-by in MENU window and
3
press the enter key (ENTER).
Change the value from 25.0 to 10.0
4
pressing the character input key.
5
Press the menu button (MENU).
Select +/− in MENU window and
6
press the enter key(ENTER).
Press the enter key(ENTER).
7
8
Press the menu button (MENU).
Select OK in MENU window and press
9
the enter key (ENTER).
CAUTION
The settable range of chamber temperature is between -10 and +60℃.
・The lighting is available only with temperature range between +2 and +50
add-on kit (MIR-L15) is installed. In the case of temperature out of range, the lighting is not usable.
The lamp is off with the temperature out of range when the saved program is running.
In this case, temperature fluctuation may exceed ±1.5
chamber temperature.
・The temperature control range is -10∼+50
The temperature is limited to 50
this case, the temperature setting over 50
Operation
o
C when the saved program includes temperature setting over 50
o
C is indicated by blinking.
Key
----
Top screen is displayed.
MENU
MENU window is displayed.
Stand-by Setting screen is displayed and chamber
ENTER
temperature setting value is changed into white
letters.
Chamber temperature setting value 10.0 is
1 0 0
displayed.
MENU
MENU window is displayed.
Chamber temperature setting value -10.0 is
ENTER
displayed.
Next item(Light Step)is displayed in white letters.
ENTER
MENU
MENU window is displayed.
The set value is memorized and top screen is
ENTER
displayed.
o
C or it may take longer time to pull down the
o
C when the optional inner door (MIR-55ID) is installed.
18
LCD Display
o
C when the optional light
o
C. In