Patient Screen (continued)
Maintaining the Patient Database
CAUTION
NOTE:
NOTE:
NOTE:
1-74
To maintain optimum system performance and to safeguard
patient data, keep the total number of patients in the database
below 1,000.
To reduce the total number of patients in the database, perform
the following procedure.
1. Before starting EZBackup, select "Unlock All" on the
Utility -> Admin -> Logon screen.
Prepare the unformatted CD-R or DVD-R before EZBackup.
Pre-formatted CD-R or DVD-R cannot be used for
EZBackup.
2. First perform EZBackup and then Backup (Patient Archive
and Report Archive).
3. Go to the patient screen, select the patients/exams to
delete. Select "Delete" to delete the selected data.
Removing image data with the "EZMove" function does not
reduce the patient number in the database.
Ensure that all patients are exported or backed up BEFORE
deleting them.
We recommend attaching the patient list to the EZBackup
media. Insert the media and select DICOM CD Read for the
dataflow (if you use a USB drive, select DICOM USB Drive
Read). Press the right Set key to select the patient. Press the
left Set key to bring up the print menu. Select Print and press
the right Set key to print the patient list.