Getting Started
Crash Recovery Instructions
Unexpected Power Loss
NOTE:
1-26
In case of a system crash, power cycle the system. Upon
boot-up, all images and measurements, except for generic
worksheets, are preserved in the system. When the system
returns, the system alerts you that unsaved images are still in
the system from the previous patient. Respond to the prompt to
continue the current patient. Check that all images and
measurements have been preserved in the system. Then
resume the exam.
If the system fails to respond to your commands within a typical
period of time, you need to manually reset the system. Simply
hold down the power switch to initiate a normal power down
sequence. After the system has completely shut down (power
switch amber), restart the system using the standard power-up
sequence. All images and measurements, except for generic
worksheets, are preserved in the system. When the system has
fully powered up, the system alerts you that unsaved images are
still in the system from the previous patient. Respond to the
prompt to continue the current patient. Check that all images
and measurements have been preserved in the system. If you
do not have any images on the clipboard, the patient must be
retrieved from the database. Then resume the exam.
In case of unexpected power loss,
1. Check the system On/Off LED color (green is on, amber is
off).
2. If the LCD monitor shuts down along with the system, then
the system has detected a power failure. It will then activate
the extended power supply (EPS). The EPS permits us to
maintain the integrity of the scanner database. The system
On/Off switch will be green in this situation.
3. Wait about 2 minutes after such an incident and then turn
the circuit breaker off (rear panel of the LOGIQ E9) or
unplug the scanner.
4. Wait approximately one minute. The system On/Off switch
should be dark.
5. Turn on the LOGIQ E9 by either turning on the circuit
breaker or plugging in the scanner.
6. Turn on the scanner by pressing the system On/Off switch.
The LOGIQ E9 should be plugged into the regular power
outlet and not the emergency power outlet.