EMC Performance (continued)
CAUTION
Table 4-4:
Portable and mobile radio communications equipment distance
Frequency Range:
Calculation Method:
Where: d= separation distance in meters, P = rated power of the transmitter, V
conducted RF, E
= compliance value for radiated RF
1
If the maximum
transmitter power in
watts is rated
5
20
100
LOGIQ E9
User Guide
Direction 5454884-100 English Rev. 1
Do not use devices which intentionally transmit RF signals
(cellular phones, transceivers, or radio controlled products),
other than those supplied by GE (wireless microphone,
broadband over power lines, for example) unless intended for
use with this system, in the vicinity of this equipment as it may
cause performance outside the published specifications.
Keep power to these devices turned off when near this
equipment.
Medical staff in charge of this equipment is required to instruct
technicians, patients and other people who may be around this
equipment to fully comply with the above regulation.
Portable and mobile radio communications equipment (e.g.
two-way radio, cellular/cordless telephones, wireless computer
networks) should be used no closer to any part of this system,
including cables, than determined according to the following
method:
requirements
150 kHz - 80 MHz
d=[3.5/V
] square root
1
of P
The separation distance in meters should be
2.6
5.2
12.0
80 MHz - 800 MHz
d = [3.5/E
] square root
1
of P
=compliance value for
1
2.6
5.2
12.0
Safety Precautions
800 MHz - 2.5 GHz
d = [7/E
] square root of
1
P
5.2
10.5
24.0
4-21