Check Function Messages - ABB Endura AZ series User Manual

Integral and remote transmitter, combustion oxygen monitor
Table of Contents

Advertisement

Endura AZ series integral and remote transmitter
Combustion oxygen monitor

8.2.4 Check Function Messages

Message range: C020.041 to M002.0XX
Status
Diagnostic Message
Icon
Displayed
––C020.041––
SIMULATION
MODE
Signals and/or
diagnostics
being simulated
––C019.044––
FACTORY TEST
MODE
Set to run mode
before using in
the field
––C018.043––
DEMONSTRATION
MODE
Set to run mode
before using
in the field
––C016.045––
CURRENT OUTPUT 1
IN TEST MODE
Output under
manual control.
No action req'd
––C014.046––
CURRENT OUTPUT 2
IN TEST MODE
Output under
manual control.
No action req'd
Table 8.5 Diagnostics – Check Function Messages
86
Possible Cause(s)
Transmitter is in simulation mode.
Transmitter in special factory test
mode.
Transmitter in demonstration
mode.
Current Output 1 under manual
control.
Current Output 2 under manual
control.
8 Troubleshooting
Corrective Measure(s)
Exit simulation mode before using in
the field.
Set to normal run mode before using
in the field.
Set to normal run mode before using
in the field.
No action required.
No action required.
IM/AZ20E–EN Rev. F

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Endura az25Endura az20Endura az10

Table of Contents