Fitting An Adaptor Box To The Sensor Cable; Adaptor Box Environmental Protection - ABB CalMaster2 Instruction Manual

Tests the integrity of a flow metering system by altering a number of flow measurement parameters and measuring the values returned from the flow measuring system.
Hide thumbs Also See for CalMaster2:
Table of Contents

Advertisement

Field and Software Tools
CalMaster2 and CheckMaster

3.3.6 Fitting an Adaptor Box to the Sensor Cable

Fig. 3.11 Fitting an Adaptor Box to the Sensor Cable
To fit an adaptor box to the sensor cable:
1.
Remove the cover of the adaptor box.
Feed the sensor lead into the adaptor box A.
2.
Connect the sensor cable B.
3.
4.
If required, provide environmental protection – see Section 3.3.7, page 16.
5.
Tighten the gland and replace the covers.

3.3.7 Adaptor Box Environmental Protection

Fig. 3.12 Potting the Adaptor Box
Warning.
Potting materials are toxic – use suitable safety precautions.
Read the manufacturers' instructions carefully before preparing the potting material.
Check all connections before potting – see Section 3.3.6, page 16.
Do not overfill or allow the potting material to come into contact with 'O' rings or grooves.
16
WABC2029
B
A
ABB
Pin 1
Brown
Pin 2
Red
Pin 3
Orange
Pin 4
Yellow
Pin 5
Green
Pin 6
Blue
Pin 7
Violet
3 Installation
Belden 8777
Black
Green
Black
Red
Sleeved Grounds
Black
White
IM/VCM2 Issue 6