  • Page 2 St Neots, U.K. – Cert. No. Q5907 Stonehouse, U.K. – Cert. No. FM 21106 As a part of ABB, a world leader in process automation technology, we offer customers application expertise, service and support worldwide. EN 29001 (ISO 9001) We are committed to teamwork, high quality manufacturing, advanced technology and unrivalled service and support.
  • Page 3: Table Of Contents

    Wiring the Adaptor Socket 5 2.2.4 System Setup ....6 OPERATION ........7 3.1 Software Installation ...... 7 3.1.1 Windows 3.1 or 3.11 ..7 3.1.2 Windows 95 ...... 7 3.2 Initialising the CalMaster ....7 3.3 Battery Charging ......7...
  • Page 4: Preparation

    Unpack and visually inspect the CalMaster. To enable a MagMaster system to be tested Also packed with the CalMaster are: by the CalMaster it is necessary to initially fit a) CalMaster sensor test lead set. a small adaptor board to the MagMaster transmitter.
  • Page 5: Preparing The Magmaster Transmitter For Verification

    2 INSTALLATION… 2.2 Preparing the MagMaster Transmitter for Verification 2.2.1 Removing the Covers Remove Protection Cover Slide Down Pull Out Slightly... Slacken ...and Captive Slide Off Screws Fig. 2.2 Access to fit Adaptor Board...
  • Page 6: Fitting The Adaptor Board

    …2 INSTALLATION 2.2.2 Fitting the Adaptor Board side and arrange the adaptor board so (Fig. 2.3 and 2.4) that the extended pins fit into the a) Turn off the power to the MagMaster. terminals previously occupied by the sensor wiring – see Fig.2.4. b) Record the position of the sensor cable wiring and remove the wiring from the d) Fit and tighten the securing screw to the...
  • Page 7: Wiring The Adaptor Socket

    2 Insert the sensor cable wiring into the normal working or the MagMaster can be socket as shown in Fig 2.5, and tighten tested by the CalMaster. the plug screws with the screwdriver provided. Caution. Ensure that no bare wires touch each other;...
  • Page 8: System Setup

    …2 INSTALLATION 2.2.4 System Setup (Fig. 2.6) a) Connect lead 3 of the CalMaster to the MagMaster 9-pin D-Type socket. The three small probes should not be connected until Warning. Care is needed when requested to do so by the CalMaster software.
  • Page 9: Operation

    'ABB Applications' folder and double click on the 'CalMaster' icon. The program Insert the first CalMaster floppy disk into the automatically finds the MagMaster and opens computer disk drive and proceed as follows: up to the Main 'ABB' Screen.
  • Page 10 NOTES...
  • Page 11 PRODUCTS & CUSTOMER SUPPORT A Comprehensive Product Range Customer Support ABB Kent-Taylor provides a comprehensive after sales Analytical Instrumentation service via a Worldwide Service Organization. Contact • Transmitters one of the following offices for details on your nearest On-line pH, conductivity, and dissolved oxygen Service and Repair Centre.