ABB CalMaster2 Instruction Manual page 23

Tests the integrity of a flow metering system by altering a number of flow measurement parameters and measuring the values returned from the flow measuring system.
Hide thumbs Also See for CalMaster2:
Table of Contents

Advertisement

Field and Software Tools
CalMaster2 and CheckMaster
8.
Select 'Yes' or 'No'. A prompt appears to test the pulse outputs:
Connection
Check pulse outputs
Yes
9.
Select 'Yes' or 'No'. A series of self-calibration screens appear. These ensure that the
CalMaster2's reference calibration, battery and data integrity are acceptable. The
Connection complete screen appears:
Connection
Connection complete
Ok
10. Select 'Ok'. CalMaster2 starts the AutoTest sequence:
Note. The first time CalMaster2 tests a flow meter that does not have an associated
fingerprint file, it must find a valid test that can be used as a reference. A screen appears
stating that CalMaster2 is searching for a reference test.
For each test there are three possible results:
Test Passed:
No action is necessary.
Test Failed:
Electrode1 impedance
Ok
Note. This screen appears for 10 seconds before proceeding automatically to
the next test.
1.
2.
Test Marginal:
Marginal (2.006)
Ok
Note. This screen appears for 10 seconds before proceeding automatically to
the next test.
1.
2.
IM/VCM2 Issue 6
No
AutoTest
Fail (0.358)
Retry
Select 'Retry'.
If the test fails again, either press 'Retry' to test again or 'Ok' to continue
to the next test.
AutoTest
I.out span
Retry
Select 'Retry'.
If the test fails again, either press 'Retry' to test again or 'Ok' to continue
to the next test.
4 Testing
21