Power On ....................20 1.11.2 Power Off ....................20 1.11.3 Power up from Standby Mode ..............21 1.12 Druck DPI 620 Genii, Modes ................22 1.12.1 Dashboard navigation ................23 1.12.2 Set Date, Time and Language ..............24 1.12.3 Themes ....................24 1.12.4...
Page 3
User Manual Druck DPI620 Genii To View Alarms select: ....................25 1.13 Software and Firmware Upgrades ..............26 1.13.1 Viewing Software Revision ..............26 1.13.2 Upgrading the Software ................26 1.14 Maintenance ....................31 1.14.1 Cleaning ....................31 1.15 Instrument Return ..................32 1.15.1...
Page 4
User Manual Druck DPI620 Genii Example Procedure: Measure AC Voltage (CH1), 0 to 20 Vrms Only ..... 52 Example Procedure: Measure AC Voltage (CH1) with the AC Probe ....53 Example Procedure: Source DC Voltage (CH1) ..........54 Example Procedure: Measure or Source Current with 24V Loop ....55 2.10...
Page 5
User Manual Druck DPI620 Genii Chart View ....................... 88 File Management .................... 89 4.5.1 Transfer ....................89 4.5.2 Erase ......................90 4.5.3 Memory Status ..................90 Data Format ....................91 Documentation ....................... 92 Analysis ......................92 Set-up ......................94 5.2.1 Define the Reference Channel ..............
Page 6
User Manual Druck DPI620 Genii 6.10 HART® Toolbar ....................110 6.11 Data Display ....................111 6.12 Editing Values ....................112 6.13 Executing Methods ..................114 6.13.1 Method Example - Self-test ..............115 6.13.2 Method Example - Analog Trim ............116 6.14...
Page 7
User Manual Druck DPI620 Genii Foundation™ Fieldbus ................... 130 Introduction ....................130 Start up ......................130 FOUNDATION™ Fieldbus Toolbar ..............133 Scanning For Devices ..................135 8.4.1 Context Sensitive Menu ................ 137 8.4.2 Troubleshooting ..................139 Device Focus View ..................140 The Navigation Menu Tree ................
Page 8
User Manual Druck DPI620 Genii Procedures (CH1/CH2): Current (source) ............. 166 Procedures (CH1/CH2): DC mV/Volts (measure) .......... 168 Procedures (CH1): DC mV/Volts (source) ............. 170 Procedures (CH1): Frequency (measure/source) ......... 172 Procedures (CH1): Frequency Amplitude (source) ........178 Procedures (CH1): Resistance measure) ............180 Procedures (CH1): True Ohms (measure) .............
The touch-screen displays up to six different parameters. This version of the manual is applicable to software revisions DK420 v2.01 and above. Equipment in the Box The following items are supplied with the Druck DPI 620 Genii: • DC power supply • Li-polymer battery •...
(MC 620) or a Pressure Station (PV 62X) to enhance the pressure measurement functionality. • Pressure Stations, PV 62X, if the Druck DPI 620 Genii is installed in a Pressure Station, it becomes a fully integrated pressure calibrator.
User Manual Druck DPI620 Genii General Safety Precautions Read and obey all the operator's local Health and Safety regulations and Safe Working Procedures or Practices. When doing a procedure or task: • Use only the approved tools, consumable materials and spares to operate and maintain the equipment.
User Manual Druck DPI620 Genii General Warnings • It is dangerous to ignore the specified limits for the instrument or its related accessories. This can cause injuries. • If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
User Manual Druck DPI620 Genii Electrical warnings • To prevent electrical shocks or damage to the instrument, do not connect more than 30V CAT I between the terminals, or between the terminals and the ground (earth). • External circuits should have appropriate insulation to the mains.
Page 14
User Manual Druck DPI620 Genii • Note that the operating and storage temperature range of the mains PSU does not match that of the DPI620. Mains PSU operating temperature range 0°C to +40°C, storage temperature range -40°C to +70°C. • The DC input to the DPI620 Genii is rated at 5V (+/-5%).
Page 15
User Manual Druck DPI620 Genii The following summary of installation and measurement overvoltage categories are derived from IEC61010-1. The overvoltage categories indicate the severity of overvoltage transients Overvoltage Description Category Overvoltage category I has the least severe overvoltage transients. Generally CAT I...
• To prevent damage to the Druck DPI 620 Genii, only use it within the specified pressure limits. • Do not exceed the maximum pressures stated in the appropriate component manual for the unit under test.
User Manual Druck DPI620 Genii Preparing the Instrument On receipt of the instrument check the contents in the box, listed in section 1.1. It is recommended to retain the box and packaging for future use. Install the Battery Remove the five Pozidriv screws (A) (Ref: Figure 1-1).
Page 18
User Manual Druck DPI620 Genii Figure 1-1 28 April 2014 K0541 issue 2 Page 18 of 200...
User Manual Druck DPI620 Genii 1.10 Charge the Battery Connect the DC power supply into the +5V DC connection on the side of the unit (Ref: Figure 1-3). The battery can also be charged using the USB connections (Ref: Figure 1-3).
STANDBY – DPI620G placed in standby mode – reduced power consumption from operating mode – recommended if unit is to be inactive for short periods. (Druck DPI 620 Genii has fast turn on from STANDBY mode). Note: SWITCH OFF can also be achieved by pressing and holding the power button until the screen is blank.
User Manual Druck DPI620 Genii 1.11.3 Power up from Standby Mode When powered-up from standby mode the instrument always opens the last screen shown before going into standby mode. Figure 1-3 28 April 2014 K0541 issue 2 Page 21 of 200...
User Manual Druck DPI620 Genii 1.12 Druck DPI 620 Genii, Modes The Druck DPI 620 Genii can be used as follows: • Calibrator (with independent functions on each of six channels). - Data logging capabilities - Documenting capabilities • HART® Communicator.
User Manual Druck DPI620 Genii 1.12.1 Dashboard navigation The Dashboard is navigated by swiping a finger from top to bottom while touching the screen. Functions screens are navigated by swiping a finger from right to left while touching the screen.
Druck DPI 620 Genii Manual Select the Help icon on the Dashboard to access the manual. All the information required to operate the Druck DPI 620 Genii, is in the Help section of the Dashboard which is accessed by selecting: DASHBOARD >>...
User Manual Druck DPI620 Genii To View Alarms select: DASHBOARD >> STATUS >> ALARM STATE Figure 1-6 Alarm State Selecting the Alarm will clear the indication until the next power down. 28 April 2014 K0541 issue 2 Page 25 of 200...
Druck DPI620 Genii 1.13 Software and Firmware Upgrades 1.13.1 Viewing Software Revision The software revisions running on the Druck DPI 620 Genii can be viewed by selecting: DASHBOARD >> STATUS >> SOFTWARE BUILD Note: If the software revision number is highlighted red then an upgrade is available.
Page 27
User Manual Druck DPI620 Genii • Upgrade the Application software and SDC625 Application. 1. Copy the ‘AMC’ application folder into the root of a USB flash memory device. 2. Put the USB flash memory device in the USB type A connector.
Page 28
User Manual Druck DPI620 Genii • Upgrade the HART processor Application and Boot Loader 1. Create a folder named ‘HART’ in the root of a USB flash memory device. 2. Copy the files ‘DK416.s19’ and ‘DK417.s19’ into the ‘HART’ folder.
Page 29
User Manual Druck DPI620 Genii • Upgrade the HART Device Library By default the HART device library is stored on the micro SD Card. 1. Set the DPI620 Genii USB client port to Storage Device mode by selecting: DASHBOARD >>...
Page 30
User Manual Druck DPI620 Genii Figure 1-7 Hart DD Directory Structure Note: • If a mistake is made during upgrade and there are no files to upload, follow the on-screen instructions and complete the procedure. • When an upgrade completes normally, the initial operation of the touch screen may be slower (a period of approximately 30 seconds).
User Manual Druck DPI620 Genii 1.14 Maintenance The DPI620 Genii instrument contains no user serviceable parts and should be returned to a GE service center for repair. 1.14.1 Cleaning CAUTION Do not use solvents or abrasive materials. Clean the case and display with a lint free cloth and a weak detergent solution.
GE Service Centre or Approved Service Agent. Contact the GE Service Centre, either by 'phone, fax or e-mail to obtain a Returned Material Authorization (RMA) number, providing the following information: Product (i.e. Druck DPI 620 Genii) Serial number Details of defect/work to be undertaken Operating conditions...
GE Service Centre or Approved Service Agent. Contact the GE Service Centre, either by 'phone, fax or e-mail to obtain a Returned Material Authorization (RMA) number, providing the following information: Product (i.e. Druck DPI 620 Genii) Serial number Details of defect/work to be undertaken Operating conditions...
User Manual Druck DPI620 Genii 1.15.3 Instrument Disposal in the European Union Do not dispose of this product or its battery as household waste. Use an approved organization that collects and/or recycles the applicable item. For more information contact • GE Sensing customer service department •...
User Manual Druck DPI620 Genii 1.16 Packaging for Storage or Transportation To store the unit or to return the unit for calibration or repair carry out the following procedures: Pack the Instrument. To return the instrument for calibration or repair, complete the return goods procedure (Ref: 1.14).
User Manual Druck DPI620 Genii Electrical Operations Basic Calibrator Operation Select: DASHBOARD >> CALIBRATOR Select the channel by performing the following tasks: • Swipe to the TASK MENU by swiping the display from right to left. Figure 2-1 Task Menu...
User Manual Druck DPI620 Genii 2.1.1 Saving Tasks At any point within the TASK MENU the currently active tasks can be saved to FAVOURITES (refer to section 2.1.3) by selecting Save Task. Note: Saved Function is what is currently active in the calibrator window.
User Manual Druck DPI620 Genii 2.1.2 Electrical Tasks Select ELECTRICAL TASKS from the TASK MENU. • This will allow the user to select from commonly used combinations of electrical functions Figure 2-2 Electrical Tasks Select the required function by selecting on either the appropriate text or diagram.
Page 39
User Manual Druck DPI620 Genii Functions can be copied to FAVOURITES (refer to section 2.1.3) by selecting as shown in Figure 2-3 and selecting Copy Task. Figure 2-3 Selected Task If the required task is not available as a Default, a new task should be created using CUSTOM TASK.
User Manual Druck DPI620 Genii 2.1.3 Favourites Selecting FAVOURITES from the TASK MENU allows selection of all SAVED and COPIED tasks. Figure 2-4 Favourites Select the required function by selecting on either the appropriate text or diagram. The DPI620G will set the functions and return to the CALIBRATOR screen.
User Manual Druck DPI620 Genii 2.1.4 Custom Task Select CUSTOM TASK from the TASK MENU. • This will allow the user to set up Channels 1 & 2 in addition to the Pressure Channels, USB (IDOS) and Communications (HART or Foundation Fieldbus).
Page 42
User Manual Druck DPI620 Genii is used for Hart® and FOUNDATION™ Fieldbus. (Refer to sections 6, 7 and 8). Figure 2-6 Channel Settings Menu 28 April 2014 K0541 issue 2 Page 42 of 200...
Page 43
User Manual Druck DPI620 Genii Setup a channel for measurement. • DIRECTION selects source or measure for the selected function • FUNCTION selects the function required (eg Current or Voltage). For more options scroll down the menu by swiping the display from bottom to top.
User Manual Druck DPI620 Genii Set the Function Utility Options For each function only one utility may be active. Not all source and measure functions have associated utilities. For all options the button resets the additional readings. 2.2.1 Max/Min This utility is only available with measure functions.
User Manual Druck DPI620 Genii 2.2.2 Switch Test This utility is available with measure or source functions. The additional values displayed show signal values (measure or source) when the instrument detects a switch opening and closing. The difference between the two values is displayed as hysteresis value for the switch.
User Manual Druck DPI620 Genii 2.2.3 Relief Valve This utility is only available with measure functions. This utility tests circuits or mechanisms that have a cut-out response when an input reaches a defined threshold value. The utility allows the user to select a mode of operation which can be rising or falling.
Page 47
User Manual Druck DPI620 Genii Rising Falling Figure 2-10 Relief Value Utility 28 April 2014 K0541 issue 2 Page 47 of 200...
User Manual Druck DPI620 Genii Measurement Display Options There are 2 display views in the CALIBRATOR screen when multiple channels are in use: • Figure 2-11 displays a reduced view of all the selected channels. Figure 2-11 Calibrator Window – Reduced View...
Page 49
User Manual Druck DPI620 Genii • Figure 2-12 displays an expanded view of the selected channel and minimizes the remaining channels. Figure 2-12 Calibrator Window - Expanded View The displays options can be changed by pressing the channel the user wants to display in expanded view.
User Manual Druck DPI620 Genii Example Procedure: Measure or Source Current Figure 2-13 shows CH1 set-up to measure or source a current with external loop power. Note: When using the CH2 connectors, set CH 2 to measure or source these ranges with internal or external loop power.
User Manual Druck DPI620 Genii Example Procedure: Measure DC Voltage Figure 2-14 shows CH1 set-up to measure a DC voltage (0 to 30 V) or DC mV (0 to 2000 mV) Note: When using the CH2 connectors, set-up CH2 to measure these range.
User Manual Druck DPI620 Genii Example Procedure: Measure AC Voltage (CH1), 0 to 20 Vrms Only WARNING To prevent electrical shocks, use only the GE specified AC probe (Part: IO620-AC) to measure AC voltages that are more than 20 Vrms (maximum: 300 Vrms). Refer to Section 2.7.
User Manual Druck DPI620 Genii Example Procedure: Measure AC Voltage (CH1) with the AC Probe WARNING To prevent electrical shocks, use only the GE specified AC probe (Part: IO620-AC) to measure AC voltages that are more than 20 Vrms (maximum: 300 Vrms). Attach it to the specified connections only.
User Manual Druck DPI620 Genii Example Procedure: Source DC Voltage (CH1) Figure 2-17 shows CH1 Set-up to source a DC voltage on CH1 (0 to 20V). Figure 2-17 Source Voltage on CH1. Range 0 to 20V. 1. Set the applicable software options.
User Manual Druck DPI620 Genii Example Procedure: Measure or Source Current with 24V Loop Figure 2-19 and Figure 2-18 show CH2 set-up to measure (±55mA) or source (0 to 24mA) a current with internal loop power (Selectable to 24 V or 28V).
Page 56
User Manual Druck DPI620 Genii Set the applicable software options. 1. Complete the electrical connections and continue with the measure or source operation. 2. Source only (Automation): Set the applicable output value. Note: The current limit of the Loop power is 30mA.
User Manual Druck DPI620 Genii 2.10 Example Procedure: Measure or Source Frequency Signals Figure 2-20 and Figure 2-21 show CH1 set-up to measure or source a frequency. The units could be Hz, kHz or counts (cpm or cph). Figure 2-20 Example A - Measure Frequency on...
Page 58
User Manual Druck DPI620 Genii Example A Example B Measure frequency on Source frequency on CH1 Range: 0 to 50 kHz Range: 0 to 50 kHz Waveform: Square Trigger level: 2.5V Amplitude:5.0 V 1. Set the applicable 1. Set the applicable software...
User Manual Druck DPI620 Genii 2.11 Example Procedure: Measure/Simulate a Resistance Temperature Detector (RTD) Figure 2-22 and Figure 2-23 show CH1 set-up to measure or simulate an RTD. A 4-wire configuration gives the best accuracy; a 2-wire configuration has the lowest accuracy (4- wire RTD shown).
Page 60
User Manual Druck DPI620 Genii Note: To measure or simulate resistance Ω, select the Resistance function (Range 0 to 4000 Ω). 1. Set the applicable software options. 2. Complete the electrical connections. 3. If necessary change the RTD Type. SETTINGS >> RTD TYPE...
User Manual Druck DPI620 Genii 2.12 Example Procedure: Measure or Simulate a Thermocouple (TC) Figure 2-24 and Figure 2-25 show CH1 set-up to measure or simulate a TC temperature. To measure or simulate TC millivolts, set the TC mV function.
Page 62
User Manual Druck DPI620 Genii Note: To measure or simulate TC millivolts, set the TC mV function 1. Set the applicable software options. 2. Complete the electrical connections. 3. If necessary change the thermocouple type. SETTINGS >> TC TYPE 4. Set CJ Compensation mode.
User Manual Druck DPI620 Genii 2.13 Example Procedure: Switch Test CH1, P1, P2 and IDOS functions use the CH2 switch connections. CH2 functions use the CH1 switch connections. Switch operation When setting the Switch Test utility on one channel, the software automatically sets-up the other channel for the switch connections.
Page 64
User Manual Druck DPI620 Genii Figure 2-26 CH1 Thermocouple Source CH2 Switch Test Figure 2-26 shows a thermocouple switch test. 1. Set the applicable software options: • The TC is set to source a temperature • The UTILITY is set to SWITCH TEST. The AUTOMATION is set to RAMP 2.
Page 65
User Manual Druck DPI620 Genii 6. Use to start the “Ramp” cycle. 7. Use to stop the “Ramp” cycle. 8. If necessary, supply the output values in the opposite direction until the switch changes condition again. 9. The display will show the following: •...
IDOS instrument. Before using an IDOS module, (Ref: User Manual: K0378, Druck IDOS UPM). Note: To attach an IDOS module to the Druck DPI 620 Genii calibrator use an IO620-IDOS-USB adaptor. 28 April 2014...
User Manual Druck DPI620 Genii 2.14.1 IDOS Option Assembly Instructions Step Procedure Attach one end of the adaptor IO620- IDOS-USB to the IDOS module Push the Type A end of USB Cable into the USB socket on the instrument and...
User Manual Druck DPI620 Genii 2.14.2 IDOS Function Procedures Set the calibrator functions for use on the display. This includes: • IDOS function (to measure the pressure) • If necessary, change the Units for the function • If necessary, set a Utility for the function:...
User Manual Druck DPI620 Genii 2.15 Error Indications Under range: The display shows this symbol for this condition: <<<<< Reading < 102% Negative Full Scale Over range: The display shows this symbol for this condition: >>>>> Reading > 102% Positive Full Scale If the display shows <<<<...
User Manual Druck DPI620 Genii Pressure Indicator Operation (MC620) This section gives examples of how to connect and use the instrument to measure pressure with the module carrier (MC 620) and the applicable pressure modules (PM 620). Figure 3-1 MC620G...
Pressure and electrical connections for a pressure module (PM 620). These are self- sealing pressure connections. Two screws to attach the calibrator (Druck DPI 620 Genii). Electrical connections for the calibrator (Druck DPI 620 Genii). Pressure module (PM 620) with a pressure connection.
Page 72
To prevent damage to the PM620 module, only use it within the specified pressure limit on the label. When the items are attached to the Druck DPI 620 Genii it is a fully integrated pressure indicator, measuring pneumatic or hydraulic pressure.
User Manual Druck DPI620 Genii 3.1.1 Assembly Instructions Step Procedure Align the two slots (a) on the calibrator with the two posts (b) on the module carrier. When the posts are fully engaged in the slots, tighten the two screws (2) hand-tight.
User Manual Druck DPI620 Genii Pressure Connections WARNING Pressurized gases and fluids are dangerous. Before attaching or disconnecting pressure equipment, safely release all the pressure. The pressure ports for external equipment use “Quick fit” pressure adaptors. (Refer to Figure 3-4).
Page 75
User Manual Druck DPI620 Genii Step Procedure (refer to Figure 3-5) Remove the adaptor from the pressure port. Use an applicable seal for the pressure connection: NPT type: Use an applicable sealant on the thread. BSP (parallel) type: Use the applicable bonded seal at the bottom.
User Manual Druck DPI620 Genii Procedure Overview Figure 3-6 Task Menu When PM620 Pressure modules are fitted, PRESSURE TASKS, are available in the TASK MENU. Refer to Section 2.1 Basic Calibrator Operation for details. 28 April 2014 K0541 issue 2...
Page 77
User Manual Druck DPI620 Genii Figure 3-7 Pressure Tasks Select the required function by selecting on either the appropriate text or diagram. The DPI620G will set the functions and return to the CALIBRATOR screen. Pressure Functions can also be selected through the CUSTOM TASK function.
Page 78
User Manual Druck DPI620 Genii If required, change the Units or the function. If necessary, set a Utility for the function: • Max/Min/Avg • Switch Test • Relief valve • Leak Test Figure 3-8 Channel Settings Note: UNITS and UTLILITIES are accessed through selecting the function through CUSTOM TASK.
User Manual Druck DPI620 Genii Set up a Leak Test This utility is only available in Pressure Measurement modes. This utility provides a test to calculate the leak of a system. Figure 3-9 Leak Test Example 28 April 2014 K0541 issue 2...
Page 80
User Manual Druck DPI620 Genii To configure leak test: Set the Utility to Leak Test. Select: SETTINGS >> LEAK TEST Set the following periods WAIT TIME: The time before the test starts in hours:minutes:seconds (hh:mm:ss). TEST TIME: The period of the leak test in...
User Manual Druck DPI620 Genii Set the Pressure Module to Zero SETTINGS >> ZERO >> ZERO Use this option to write a new zero pressure value to the pressure module in use. The sensor adjustment is permitted if it obeys the condition that follows: •...
User Manual Druck DPI620 Genii Error Indications Under range: The display shows this symbol for this condition: <<<<< Reading < 110% Negative Full Scale Over range: The display shows this symbol for this condition: >>>>> Reading > 100% Positive Full Scale If the display shows <<<<...
User Manual Druck DPI620 Genii Data Logging Operation Select the DATA LOGGING option on the Dashboard. The Data Logging function records instrument readings so they can be reviewed or analysed. Figure 4-1 Data Logging The data file can be reviewed by using the following: •...
Page 84
User Manual Druck DPI620 Genii This chapter describes how to use the Data Logging function to log data to a file. In Data Logging mode the display data from all active channels is stored at each data point. The data can be stored: •...
User Manual Druck DPI620 Genii Set-up Before starting, set all channels to the correct functions. (Refer to section 2.1). To access the Data logging function do the following: DASHBOARD >> DATA LOGGING >> SETUP Figure 4-2 Datalogging Setup 28 April 2014...
Page 86
User Manual Druck DPI620 Genii • STORAGE AREA Used to set Internal or SD card storage. Only the SD card can be read when connected to a PC. • FILENAME Enter the filename (10 characters maximum) • TRIGGER Select one of the following: ...
User Manual Druck DPI620 Genii Operation In periodic mode, to begin data logging tap ‘Start logging’ button. In Key press mode, a data point is taken every time the user taps the log button. To stop Data Logging select .
User Manual Druck DPI620 Genii Chart View DASHBOARD >> DATA LOGGING >> RECALL To view a data file on a chart do the following: Tap the Filename button to display the list of data files. Select the file to be displayed.
User Manual Druck DPI620 Genii File Management The data log file management options are as follows: • TRANSFER : Upload data log files to another computer. • ERASE: Delete data log files. • MEMORY STATUS: Displays amount of free memory.
User Manual Druck DPI620 Genii 4.5.2 Erase The Erase options are as follows: • ERASE ONE FILE: Select file and tap tick bottom right on the screen to erase. • CLEAR INTERNAL: Clears all internal files. 4.5.3 Memory Status The MEMORY STATUS button will show the amount of available memory in the areas that follow: •...
User Manual Druck DPI620 Genii Data Format The data files are produced in a Comma Separated Variable (csv) format (refer to Figure 4-4).. This allows the data to be imported into a spreadsheet (e.g. Microsoft® Excel). The first section of the data file contains the...
User Manual Druck DPI620 Genii Documentation This chapter describes the Documenting functions available with the Druck DPI 620 Genii calibrator and are as follows: • ANALYSIS • RUN PROCEDURE Analysis The Analysis function takes readings from two or more DPI 620 channels to calibrate the transfer characteristic of the device being tested.
Page 93
User Manual Druck DPI620 Genii A second input channel can also be used, to calculate the transfer characteristic between three points in the signal path and may be calibrated at the same time, as in the example that follows. •...
User Manual Druck DPI620 Genii Set-up Set the Druck DPI 620 Genii channels in the Calibrator function. (Refer to section 2.1). Connect the calibrator to the device under test. Enter the Documenting function. DASHBOARD >> DOCUMENTING Tap the ANALYSIS button.
User Manual Druck DPI620 Genii 5.2.1 Define the Reference Channel Tap the channel button that is to be used as the Reference channel for the analysis. Figure 5-1 Select Reference Channel Set the channel type to Reference. All other channel settings for that channel are cancelled.
User Manual Druck DPI620 Genii 5.2.2 Define each Input Channel Tap each Input channel button to set the Input options. Figure 5-2 Select Input Options • SCALING - The scaling values are four set values: The maximum and minimum Reference signal values (Reference High and Reference Low).
Page 97
User Manual Druck DPI620 Genii • ERROR TYPE – The deviation from which the transfer characteristic is to be calculated. This can be one of the following: % Span - a percentage of the Input signal span. - a percentage of the Input signal reading.
User Manual Druck DPI620 Genii Analysis Function Set Input channel parameters (refer to section 5.2), and return to CHANNEL SETUP screen. Select the Start button The Analysis window displays the following: • The deviation of each Input channel from the ideal transfer characteristic.
• A copy of the 4 Sight Calibration Software. • Standard USB lead (as supplied). • A Druck DPI 620 Genii calibrator device driver available as a download from the website. 28 April 2014...
4. Use the Download button in calibrator manager to Download the file to the Druck DPI 620 Genii calibrator. A communications symbol will be displayed at the bottom of the screen.
User Manual Druck DPI620 Genii HART® Operations The Druck DPI 620 Genii can communicate with devices that use the HART® protocol as follows: • The Universal and Common Practice commands specified in HART® revision 5 to 7. • Devices that support Device Descriptions (DD).
User Manual Druck DPI620 Genii The Druck DPI 620 Genii can be used to communicate with other HART® field devices as follows: • As a Primary Master the Druck DPI620 Genii starts and controls all communications. The field device (slave) uses each command from the master device to make a change and/or send data back.
Druck DPI620 Genii HART® Connections Before setting-up the electrical connections between the HART® device and the Druck DPI 620 Genii get the correct connection scheme (Ref: DASHBOARD >> HELP). Power Supply from the Calibrator A 24 V or 28 V loop power can be supplied using the CH2 Current (24V) measure function.
User Manual Druck DPI620 Genii External Loop Power In the example that follows, there is an external power supply. Figure 6-2 Measure current on CH2 without 24 V loop power. HART® function is enabled and 250 Ω resistor is enabled.
User Manual Druck DPI620 Genii Communicator Attached to a Network In the example that follows, the calibrator connects directly to a network. There must be a 250 Ω resistor in series with the loop power supply and the HART® device.
User Manual Druck DPI620 Genii Viewing Primary Variables When connected to a HART® device, the Primary Variable value and PV units will be displayed in the channel window. Figure 6-5 Primary Variables 6.8.1 Device Polling The connected HART® device may be configured with a unique Poll Address.
User Manual Druck DPI620 Genii use a specific Poll Address. • Select “Find First Device” to search from 0 to 63 and use the first device found. • Select targeted address ranges to select from a range of devices at poll addresses.
Page 108
User Manual Druck DPI620 Genii Figure 6-6 Hart Channel Window The Main HART® SDC application will now open. The Druck DPI 620 Genii displays the HART® SDC application screen in light or dark mode. 28 April 2014 K0541 issue 2...
Page 109
User Manual Druck DPI620 Genii Figure 6-7 HART® Application Main Screen 28 April 2014 K0541 issue 2 Page 109 of 200...
User Manual Druck DPI620 Genii 6.10 HART® Toolbar Figure 6-8 HART® Toolbar On entering the HART® SDC application, the toolbar is displayed. The ICONs are greyed out when they are not active. The ICON functions are described below: OPEN NEW CONNECTION – Requires HART®...
• LRV - Lower Range Value • USL - Upper Sensing Limit • LSL - Lower Sensing Limit The data display also shows the current readings of the DPI 620 Genii instrument channels. This is used in calibration procedures. 28 April 2014 K0541 issue 2...
User Manual Druck DPI620 Genii 6.12 Editing Values Any value displayed in white/black with a [V] or [E] icon can be edited. Edit variables as follows: 1. Select the variable 2. If a selection window is opened select the variable (or select Edit button).
Page 113
User Manual Druck DPI620 Genii Note: To go back to the original value, select the Abort button in the menu bar. 7. Select Commit button in the menu bar to implement a new value 8. The yellow highlight will disappear.
Exits from the procedure. ABORT Accepts inputs and proceeds to the next step. SWITCH APP: Returns to Druck DPI 620 Genii screen (to change channel function settings without interrupting the method procedure). Note: Some methods may make the HART® device output a certain current.
User Manual Druck DPI620 Genii Some methods need values to be entered. Use the alpha/numeric keypads. A drop down menu can be used for method selection options. Some methods require input from the DPI 620 instrument channels. A drop down menu displays the channels as follows: •...
User Manual Druck DPI620 Genii 6.13.2 Method Example - Analog Trim The Druck DPI 620 Genii can perform an analogue trim on the 4 to 20 mA loop without connecting to any external reference meters. 1. Navigate to the calibration folder Figure 6-10 Calibration Folder 2.
Page 117
User Manual Druck DPI620 Genii 4. Read the CH2 value and enter this in the meter value text box using the keypad. Figure 6-11 Enter Calibration Point 5. Select SET. 6. Repeat steps 3 and 4 with 20 mA selected. This will calibrate the transmitter’s output current.
Transmitters with a non-zero poll address are in multi-drop mode and default to a fixed loop current of 4 mA. By default, the Druck DPI 620 Genii polls address 0 (zero) only. Change the Poll Address by selecting the appropriate search radio button or entering the tag name in the search field.
External HART® resistor present, check the DPI 620 resistor setting is OFF. Check that the CH 2 function is set to ‘NONE’ if the Druck DPI 620 Genii is used as a secondary master (in parallel with an external supply.
User Manual Druck DPI620 Genii HART® Offline Introduction The HART® Offline feature sits alongside the existing HART® application to provide extended offline capabilities. Typical operations include the following: • Connect to a HART enabled device to view its configuration and save to file.
User Manual Druck DPI620 Genii Start HART® Offline 1. Select the HART® channel so that it is in its expanded view (refer to Section 2.3) 2. Select HART offline button Figure 7-1 Hart Channel Window 28 April 2014 K0541 issue 2...
User Manual Druck DPI620 Genii The HART® offline feature will be presented. Figure 7-2 HART Offline Main Menu On first use, an empty menu will be presented. Once Offline Configurations have been created, they will be listed in this menu (most recent first).
User Manual Druck DPI620 Genii 7.4.1 Connected Configuration Press the New Configuration button Details of the connected device will be displayed: Figure 7-3 Device Details Press the button to continue. 28 April 2014 K0541 issue 2 Page 123 of 200...
User Manual Druck DPI620 Genii 7.4.2 Disconnected Configuration Press the New Configuration button Select from the menu the device details that you want to configure. Press the button to continue. 28 April 2014 K0541 issue 2 Page 124 of 200...
User Manual Druck DPI620 Genii Edit an Offline Configuration Having created an offline configuration (either connected or disconnected) the HART Commands menu will be displayed. Figure 7-4 Commands Read/write commands appear at the top of the list. Read only commands appear at the bottom of the list and are coloured red.
Page 126
User Manual Druck DPI620 Genii Selecting a Command will display a list of Variables relevant to the Command. Figure 7-5 Variables When connected, the Variables will be populated with values from the device. When disconnected, the variables are populated by the user.
User Manual Druck DPI620 Genii Saving the Configuration From the Commands menu, the configuration may be saved by pressing the Save Button The device name will be used as the default file name. Having saved the configuration, the main Offline menu will be presented with the new configuration file first in the list.
User Manual Druck DPI620 Genii Uploading the Configuration From the Commands menu, the configuration may be uploaded to connected device by pressing the Upload Button Working with Saved Configurations Selecting a Saved Configuration from the main Offline menu enables the user to perform the following operations: 7.8.1...
User Manual Druck DPI620 Genii 7.8.4 Delete HART Config Deletes the HART configuration file from the DPI620 Genii. Further Operations 7.9.1 Delete All Configuration Files From the main offline menu, all saved configuration files may be deleted by pressing the Delete All Button.
User Manual Druck DPI620 Genii Foundation™ Fieldbus Introduction FOUNDATION™ Fieldbus (FF) is a device application for configuration of FF enabled field devices. On-line connection is via the integrated H1 modem. Devices directly connected to an H1 field segment can be configured and supported by Start up FOUNDATION™...
Page 131
User Manual Druck DPI620 Genii To start the FOUNDATION™ Fieldbus application and connect to a device: Connect DPI620 Genii to a H1 FOUNDATION™ Fieldbus device. Figure 8-1 Example Connection Diagram Foundation Fieldbus Select the FOUNDATION™ Fieldbus channel so that it is in its expanded view (refer to Section 2.3).
Page 132
User Manual Druck DPI620 Genii Select button. Figure 8-2 Calibrator Fieldbus Channel 4. The Main FOUNDATION™ Fieldbus application will now open. Note: Updating CH1, CH2, P1, P2 tasks is only possible when the FOUNDATION™ Fieldbus application is not running – to close FOUNDATION™ Field bus application select : HOME >>...
User Manual Druck DPI620 Genii FOUNDATION™ Fieldbus Toolbar Figure 8-3 FOUNDATION™ Fieldbus Toolbar On entering the FOUNDATION™ Fieldbus application, the toolbar is displayed. The ICONs are greyed out when they are not active. The ICON functions are described below: OPEN CONNECTION – Only available when awaiting an open connection.
Page 134
User Manual Druck DPI620 Genii ABORT – Abort Update of parameters, reverts to previous values (refer to Section 8.7.3). STATUS – Provides Device profile of currently connected device (refer to Section 8.5). FUNCTION FINDER – Search FF variables and device functions.
User Manual Druck DPI620 Genii Scanning For Devices The following steps describe how to scan for FOUNDATION™ Fieldbus devices over a FOUNDATION™ Fieldbus H1 connection: Connect DPI620 Genii to H1 bus (refer to section 8.2). Select the OPEN CONNECTION Icon on the Toolbar to enter device scan screen.
Page 136
User Manual Druck DPI620 Genii Select the SCAN button The ‘scanning’ progress dialog view will open. Any devices found in the selected range will appear in the bus tree window list - All scanned devices are shown as a bold icon with an associated tag. Previous scan results are indicated in grey.
User Manual Druck DPI620 Genii Selecting any device in the search results and selecting OK will initiate connection to the ‘Device Focus view.’ To Re-scan use the context sensitive menu on the Genii Modem (Refer to section 8.4.1). 8.4.1 Context Sensitive Menu Context sensitive menus are available for each device in the scan Menu by select and hold.
Page 138
User Manual Druck DPI620 Genii Devices This provides access to the following functions: • Tag / Address Change Set Tag Set Address Clear Tag Clear Address • Boot Operational Function (BOF) class change Set Basic Set Link Master Figure 8-7 Scan Menu -...
User Manual Druck DPI620 Genii 8.4.2 Troubleshooting • If no device is found when performing scan check: Field wiring. Verify electrical segment connections are in accordance with the specific manual supplied with the field device and segment coupler / power supply.
User Manual Druck DPI620 Genii Device Focus View In this view the specific information of the device is shown. – PD tag – Device Id – Block List with Target / Actual Mode On entering ‘Device Focus view’ the software will load the Blocks of the target field device and make them available for parameterization.
Page 141
User Manual Druck DPI620 Genii Selecting the Scan button returns device scan view (reference Figure 8-5). Selecting the desired block with a single key press will open the Navigation Tree for the block (Refer to section 8.6). 28 April 2014...
User Manual Druck DPI620 Genii The Navigation Menu Tree This is the overview of the entire loaded block of the device (this is not the entire field device only one aspect of it) showing all the menus available according to the configured access level.
User Manual Druck DPI620 Genii 8.6.1 7.6.1 Block Header bar The block header bar indicates the target and actual mode of the block. Figure 8-11 Block Header Bar The Highlighted text indicates the actual mode of the device block. The highlight is shown in green if the target mode is equal to the actual mode of the device block.
User Manual Druck DPI620 Genii Functional Group View The functional groups show all the variables or settings in that menu group along with the current value. Figure 8-12 Function Group View The left side hosts the 'Variable Description Area' and access to the context based adjustment functions.
Page 145
User Manual Druck DPI620 Genii Black variables are open for editing under the appropriate access conditions, such as access code or PIN that may be required in a different Functional Group. The Navigation Tree bar shows the hierarchy of menus and groups above the current Function Group View.
User Manual Druck DPI620 Genii 8.7.1 Displaying Parameter Help • The yellow triangle in the corner of the Variable Description area indicates that help is available for that parameter. • Context menu opened up with touch and hold action. • Selecting Display Help will show the help attributes.
User Manual Druck DPI620 Genii 8.7.2 Refreshing Data When a refresh is in progress the variable description goes grey and in the right area of the Variable Editing area the pending icon appears. When the read request is fulfilled the variable description goes from grey to black again and the pending icon disappears.
User Manual Druck DPI620 Genii 8.7.3 Editing Values Values which may be editable are shown in black in the “variable editing area” of the function group view (Refer to Figure 8-12). Select the parameter to open for editing. Once Edit is complete, the variable description is highlighted in bold and the commit and cancel icons in the toolbar become active.
User Manual Druck DPI620 Genii If an invalid value is entered the variable will be shown in red and the error icon appears. 8.7.4 Methods These are viewed as a button or selection menu. In the case of a button, a momentary touch activates the execution of the associated functionality.
User Manual Druck DPI620 Genii Function Finder Function Finder is a way of searching for FF variables and device functions in the on-line device. In complex devices with multiple menus this allows the user to navigate a device without a manual, greatly simplifying the on-line experience, even with an unfamiliar device.
Page 151
User Manual Druck DPI620 Genii Figure 8-17 Function Finder 28 April 2014 K0541 issue 2 Page 151 of 200...
User Manual Druck DPI620 Genii Exporting Data to Main Genii Application The FOUNDATION™ Fieldbus application allows selected parameters to displayed in the Communications channel windows for the Calibrator screen. The selected parameters are defined by the Export List Menu. Figure 8-18 Export List...
Page 153
User Manual Druck DPI620 Genii Parameters are added to the Export List using the context menu in the Variable Description area (refer to Section 8.7) and selecting ‘Add to ‘Export List’. Note: Only Parameters that return a value can be added to the Export List.
User Manual Druck DPI620 Genii 8.9.1 Viewing Exported Variables in Channel Window Return to Main Application by Minimizing the FOUNDATION™ Fieldbus application. >> HOME Minimize Expand the FOUNDATION™ Fieldbus window and select: Settings >> PRIMARY VALUE The Export List of selected parameters will be displayed. The selected parameter will be displayed in the FOUNDATION™...
User Manual Druck DPI620 Genii 8.10 My Block My Block allows the user to create Menus of commonly used parameters for easy recall. Additional Menus can be created under ‘My Block’ using the context menu accessed with touch and hold gesture.
Page 156
User Manual Druck DPI620 Genii Parameters are added to ‘My Block’ (or the created Menu) using the context menu in the Variable Description area (refer to Section 8.7) . Figure 8-22 Adding Parameters to 'My Block' 28 April 2014 K0541 issue 2...
User Manual Druck DPI620 Genii 8.11 Application Settings Application settings are access from the FOUNDATION™ Fieldbus toolbar by selecting: SETTINGS 8.11.1 Device Library The library tab shows the Device Descriptions (DDs) that are currently on the DPI620 Genii. Allows the user to browse for a specific device to determine that there is support for it.
User Manual Druck DPI620 Genii 8.11.2 Options • Poll header every - Sets the refresh rate for the device parameters displayed in the header. • Poll all dynamic every - Sets the refresh rate for FF dynamic variables in the function group view (note this setting only becomes valid if the function group option Refresh Vars On is active) (refer to Section 8.7.2).
User Manual Druck DPI620 Genii 8.11.3 Advanced These settings are for advanced users only and it is recommended to retain Default Values. 28 April 2014 K0541 issue 2 Page 159 of 200...
User Manual Druck DPI620 Genii Calibration Procedures Note: GE Measurement and Control can provide a calibration service that is traceable to international standards. Note: GE Measurement and Control recommend returning the instrument to the manufacturer or an approved service agent for calibration. If using an alternative calibration facility, check that it uses the standards that follow.
Page 161
User Manual Druck DPI620 Genii Table 9-1 Function Calibration equipment (ppm = parts per million) Current Current (mA) calibrator. (CH1 or CH2) Accuracy - Current measure/source, refer to Table 9-2 or Table 9-3 Voltage Volts calibrator. (CH1 or CH2) Accuracy - Voltage measure/source,...
Page 162
User Manual Druck DPI620 Genii Function Calibration equipment (ppm = parts per million) Thermocouple temperature reference unit (0°C) Accuracy: 30 mK AC mV AC mV calibrator. (CH1) Accuracy - AC mV measure, Refer to Table 9-15 AC Volts AC Volts calibrator.
User Manual Druck DPI620 Genii To do a calibration on a measure or source function, use the advanced menu option. 1. Select: DASHBOARD >> ADVANCED 2. Enter the calibration PIN: 4321 3. Select the button. 4. Select: PERFORM CALIBRATION Then select a function and start the calibration: 1.
Page 164
User Manual Druck DPI620 Genii 1. Connect the applicable calibration equipment (Ref: Table 9-1). 2. Let the equipment get to a stable temperature (minimum: 5 minutes since the last power on). 28 April 2014 K0541 issue 2 Page 164 of 200...
Page 165
User Manual Druck DPI620 Genii 3. Use the calibration menu (Ref: Section 9.1) to do a three- point calibration (-FS, Zero and +FS) for each range: 20 mA and 55 mA. 4. Check the calibration is correct. • Select the applicable Current (measure) function, (Ref: Section 2).
User Manual Druck DPI620 Genii Procedures (CH1/CH2): Current (source) Do the procedure as follows: 1. Connect the applicable calibration equipment (Ref: Table 9-1): • CH1/CH2 (24 mA range): (Ref: Section 2 and Section 2.9). • CH2 (-24 mA range only): (Ref: Section 2 and Section 2.9).
Page 167
User Manual Druck DPI620 Genii 4. Check the calibration is correct: • Select the applicable Current (source) function(Ref: Section 2 and Section 2.9). • Apply the values that follow: CH1/CH2: 0.2, 6, 12, 18, 24 • Check the error is within limits (Ref: Table 9-3)
User Manual Druck DPI620 Genii Procedures (CH1/CH2): DC mV/Volts (measure) When recalibrating measure functions for this range, any adjustment will affect the corresponding source function calibration. The source function must, therefore be re- calibrated after adjusting the measure function. Do the procedure as follows: 1.
Page 169
User Manual Druck DPI620 Genii Volts (V): -30, -21, -20, -10, -5, 0 (short circuit) Volts (V): 0, 5, 10, 20, 21, 30. • Check that the error is within limits (Ref: Table 9-4 or Table 9-5). Table 9-4 Millivolts (measure) error limits...
User Manual Druck DPI620 Genii Procedures (CH1): DC mV/Volts (source) Do the procedure as follows: 1. Connect the applicable calibration equipment (Ref: Table 9-1). 2. Let the equipment get to a stable temperature (minimum: 5 minutes since the last power on).
Output = 10V Unipolar Square wave Frequency = 990 Hz Druck DPI 620 Genii: Input units = Hz Input trigger level = 5 V 4. Use the calibration menu (Ref: Section 9.1) to do a one- point calibration. 5. Check the calibration is correct.
Page 173
(minimum: 5 minutes since the last power on). 3. Set-up the equipment as follows: Frequency meter: Gate time = one second Druck DPI 620 Genii: Waveform = Square; Amplitude = 10 V Frequency = 990 Hz 4. Use the calibration menu (Ref: Section 9.1) to do a one- point calibration.
Page 174
• Frequency (measure) calibration check. Signal generator: Output = 10 V Unipolar Square wave Druck DPI 620 Genii: Input trigger level = 5 V Units: Hz or kHz as specified in 28 April 2014 K0541 issue 2 Page 174 of 200...
Page 175
Table or Table 9-9 • Frequency (source) calibration check. Frequency meter: Gate time = one second • Druck DPI 620 Genii: Units: Hz or kHz as specified in Table 9-8. 28 April 2014 K0541 issue 2 Page 175 of 200...
Page 176
User Manual Druck DPI620 Genii Table 9-8 or Table 9-9 • Select the applicable Millivolts or Voltage (measure) function (ref Section 2.10). Apply the input values: 0, 990 kHz: 10, 50 • Check the error is within limits (Ref: Table 9-8 or Table 9-9).
User Manual Druck DPI620 Genii Procedures (CH1): Frequency Amplitude (source) Do the procedure as follows: Note: The procedure that follows calibrates the “mark” value of the square wave frequency output. The “space” value is fixed and is approximately -120 mV.
Page 179
User Manual Druck DPI620 Genii 5. Check the calibration is correct. • Set-up the equipment as follows: Source frequency = 0 (For DC output); Waveform = Square • Apply the amplitude values that are applicable to the calibration (Ref: Table 9-10).
User Manual Druck DPI620 Genii Procedures (CH1): Resistance measure) Do the procedure as follows: 1. Connect the applicable calibration equipment (Ref: Table 9-1). 2. Let the equipment get to a stable temperature (minimum: 5 minutes since the last power on).
Page 181
User Manual Druck DPI620 Genii 4. Check the calibration is correct • Select the applicable Resistance (measure) function (Ref Section 2.1159). • Make a 4-wire connection to the applicable standard resistor (Ref: Table 9-11) and measure the value. • Check the error is within limits (Ref: Table 9-11)
User Manual Druck DPI620 Genii Procedures (CH1): True Ohms (measure) Do the procedure as follows: 1. Repeat procedure 9.8; in step 3 and 4 select True Ohms. 2. Check the error is within limits (Ref: Table 9-1). Table 9-12 True Ohms (measure) error limits...
User Manual Druck DPI620 Genii 9.10 Procedures (CH1): Resistance (source) Do the procedure as follows: 1. Connect the applicable calibration equipment (Ref: Table 9-1). 2. Let the equipment get to a stable temperature (minimum: 5 minutes since the last power on).
Page 184
User Manual Druck DPI620 Genii 4. Check the calibration is correct: • Select the Resistance (source) function (Ref: Section 2.11). • Apply the resistance values that are applicable to the calibration (Ref: Table 8-13). • Check the error is within limits (Ref: Table 8-13).
User Manual Druck DPI620 Genii 9.11 Procedures (CH1): TC mV (measure or source) When recalibrating measure functions for this range, any adjustment will affect the corresponding source function calibration. The source function must, therefore be re- calibrated after adjusting the measure function.
Page 186
User Manual Druck DPI620 Genii TC mV (source): -10, 0, 10, 25, 50, 100 6. Check the error is within limits (Ref: 7. Table 9-14). Table 9-14 TC mV (measure or source) error limits Calibrator uncertainty Permitted Input TC (mV)
User Manual Druck DPI620 Genii 9.12 Procedures (CH1): Cold Junction (TC method) and CJ (measure) Note: Do the TC mV (measure) calibration before the cold junction calibration. The conditions for CJ calibration are as follows: • Battery mode (with DC charger disconnected) •...
Page 188
User Manual Druck DPI620 Genii CJ (TC Method) Do the procedure as follows: 1. Connect the applicable calibration equipment (Ref: Table 9-1). 2. Set the reference unit temperature: 0°C. 3. Let the equipment get to a stable temperature (minimum: 1 hour since the last power on).
Page 189
User Manual Druck DPI620 Genii Do the procedure as follows: 1. Connect the applicable calibration equipment (Ref: Table 9-1). 2. Set-up the equipment: • Function = TC (measure) • TC Type = K Type • CJ Compensation, Mode = Automatic 3.
User Manual Druck DPI620 Genii 7. Check the calibration is correct • Select the TC (measure) function 8. Check the equipment gives a TC temperature that is the same as the temperature on the reference unit ±0.1°C (0.2°F) 9.13 Procedures (CH1): AC mV/Volts (measure) Do the procedure as follows: 1.
Page 191
User Manual Druck DPI620 Genii 4. Check the calibration is correct: • Select the applicable AC mV or AC Volts (measure) function (ref Section 2.6). 5. Apply the input values that are applicable to the Calibration. • AC mV: 10, 500, 1000, 2000 •...
User Manual Druck DPI620 Genii 9.14 Procedures: Pressure Indicator Modules (PM 620) Do the procedure as follows: 1. Assemble the pressure indicator with the necessary PM620 modules. 2. Connect the instrument to the pressure standard. 3. Let the equipment get to a stable temperature (minimum: 60 minutes since the last power on).
Page 193
User Manual Druck DPI620 Genii Table 9-17 Calibration pressures Nominal applied pressure Ranges: gage mbar (psi) -FS † Zero < 700 mbar (10.0 psi) > 700 mbar (10.0 psi) -900 (-13.1) †For a three-point calibration, do not apply more than - 90% of the specified FS for the unit.
Page 194
User Manual Druck DPI620 Genii 5. Check the calibration is correct: • Select the applicable pressure function. • Apply the following pressure values (absolute sensors). • 0, 20, 40, 60, 80, 100 (%FS). • Go back to 0 in the same steps.
User Manual Druck DPI620 Genii 9.15 Procedures: IDOS UPM (Ref: User manual IDOS UPM). When the calibration is complete, the instrument automatically sets a new calibration date in the UPM. 28 April 2014 K0541 issue 2 Page 195 of 200...
User Manual Druck DPI620 Genii General Specification 10.1 Introduction For a full specification of the Druck DPI 620 Genii calibrator and its related accessories (MC 620 carrier, PM 620 module and PV 62x pressure stations) refer to the relevant product datasheet.
Page 197
-10 to 50°C (14 to 122°F) Operating temperature -20 to 70°C (-4 to 158°F) Storage temperature IP55 (Druck DPI 620 Genii calibrator only) Ingress Protection 0 to 90% relative humidity (RH) non-condensing Humidity MIL-PRF-28800F for class 2 equipment Shock/Vibration Pollution Degree...
Page 198
User Manual Druck DPI620 Genii Note 1: The DPI 620 has been assessed to the European IEC60529 standard as having an ingress protection rating of IP55, but this is for reliability purposes and not for safety reasons. Note 2: To meet the immunity requirements of annex A of...
User Manual Druck DPI620 Genii Manufacturer 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.TestEquipmentDepot.com 28 April 2014 K0541 issue 2 Page 199 of 200...
User Manual Druck DPI620 Genii Display Icons Table 12-1 Dashboard Icons Advanced Hart® Calibrator Help Data Logging LED (shows status) Blue – Active Red – Alarm Green - Connected Touchscreen Multimeter Documenting Scope Foundation Settings Fieldbus Files Status Applications Devices...